Day-and-night glycaemic control with closed-loop insulin delivery versus conventional insulin pump therapy in free-living adults with well controlled type 1 diabetes: an open-label, randomised, crossover study
Круглосуточный контроль гликемии при замкнутой доставке инсулина по сравнению с обычной помповой инсулинотерапией у взрослых с хорошо контролируемым сахарным диабетом 1 типа в условиях повседневной жизни: открытое рандомизированное перекрёстное исследование
2017-01-13
SCID: 54.1/2bx2ycud
Discuss with AI
continuous glucose monitoringhybrid closed-loop insulin deliverymodel-predictive controltime in target rangetype 1 diabetes
Figures from the paper
Abstract (AI)
Background Tight control of blood glucose concentration in people with type 1 diabetes predisposes to hypoglycaemia. We aimed to investigate whether day-and-night hybrid closed-loop insulin delivery can improve glucose control while alleviating the risk of hypoglycaemia in adults with HbA 1c below 7·5% (58 mmol/mol). Methods In this open-label, randomised, crossover study, we recruited adults (aged ≥18 years) with type 1 diabetes and HbA 1c below 7·5% from Addenbrooke's Hospital (Cambridge, UK) and Medical University of Graz (Graz, Austria). After a 2–4 week run-in period, participants were randomly assigned (1:1), using web-based randomly permuted blocks of four, to receive insulin via the day-and-night hybrid closed-loop system or usual pump therapy for 4 weeks, followed by a 2–4 week washout period and then the other intervention for 4 weeks. Treatment interventions were unsupervised and done under free-living conditions. During the closed-loop period, a model-predictive control algorithm directed insulin delivery, and prandial insulin delivery was calculated with a standard bolus wizard. The primary outcome was the proportion of time when sensor glucose concentration was in target range (3·9–10·0 mmol/L) over the 4 week study period. Analyses were by intention to treat. This study is registered with ClinicalTrials.gov, number NCT02727231, and is completed. Findings Between March 21 and June 24, 2016, we recruited 31 participants, of whom 29 were randomised. One participant withdrew during the first closed-loop period because of dissatisfaction with study devices and glucose control. The proportion of time when sensor glucose concentration was in target range was 10·5 percentage points higher (95% CI 7·6–13·4; p<0·0001) during closed-loop delivery compared with usual pump therapy (65·6% [SD 8·1] when participants used usual pump therapy vs 76·2% [6·4] when they used closed-loop). Compared with usual pump therapy, closed-loop delivery also reduced the proportion of time spent in hypoglycaemia: the proportion of time with glucose concentration below 3·5 mmol/L was reduced by 65% (53–74, p<0·0001) and below 2·8 mmol/L by 76% (59–86, p<0·0001). No episodes of serious hypoglycaemia or other serious adverse events occurred. Interpretation Use of day-and-night hybrid closed-loop insulin delivery under unsupervised, free-living conditions for 4 weeks in adults with type 1 diabetes and HbA 1c below 7·5% is safe and well tolerated, improves glucose control, and reduces hypoglycaemia burden. Larger and longer studies are warranted. Funding Swiss National Science Foundation (P1BEP3_165297), JDRF, UK National Institute for Health Research Cambridge Biomedical Research Centre, and Wellcome Strategic Award (100574/Z/12/Z).
Key Findings
1
Closed-loop therapy was designed to improve glucose control while reducing hypoglycaemia risk, although the provided abstract excerpt does not report all secondary outcomes.
2
In adults with well-controlled type 1 diabetes, hybrid closed-loop delivery increased time in target glucose range by 10.5 percentage points versus usual pump therapy.
3
One participant withdrew during the first closed-loop period because of dissatisfaction with the study devices and glucose control.
4
Participants spent 65.6% of time in target range with usual pump therapy compared with 76.2% using closed-loop delivery.
5
The randomized crossover study evaluated unsupervised, day-and-night closed-loop insulin delivery under free-living conditions over four-week treatment periods.
Research Object
Adults with well-controlled type 1 diabetes using day-and-night hybrid closed-loop or conventional insulin pump therapy under free-living conditions
Research Subject
Glycaemic control and hypoglycaemia risk, including time in target glucose range, during closed-loop versus conventional pump insulin delivery
Publication Details
Publication Date
2017-01-13
Journal
Publisher
ISSN
Cited by
156
Open access PDF
Access Type
Author Information
Download PDF
Subscribe to digest