Determination of Chlormequat in Fruit Samples by Liquid Chromatography-Electrospray-Mass Spectrometry/Mass Spectrometry

Определение хлормеквата в образцах фруктов методом жидкостной хроматографии с электроспрей-ионизацией и тандемной масс-спектрометрией
R. Castro, Encarnación Moyano, Maria Teresa Galceran
2001-11-01

chlormequat determinationfruit samplesion-pair chromatographyliquid chromatography-tandem mass spectrometrysolid-phase extraction
An ion-pair liquid chromatography-mass spectrometry/mass spectrometry method was developed for the determination of chlormequat in fruit samples. A solid-phase extraction cleanup procedure with C18 cartridges was used. Sample preparation was simple and the achieved detection limit (0.03 mg/kg) and quantitation limit (0.08 mg/kg) were below the maximum residue levels legislated by the European Union. The chromatographic separation was performed by using a C8 column and heptafluorobutyric acid as ion pair reagent. The detection was conducted with an electrospray source and an ion trap as a mass analyzer. The reproducibility of the method gave good run-to-run (8-9%) and day-to-day (9-13%) precision values, and its applicability to the determination of chlormequat in pears and grapes, purées, and pear, apple, and grape juices was demonstrated.
1
An ion-pair liquid chromatography–electrospray tandem mass spectrometry method was developed to determine chlormequat in fruit samples.
2
C18 solid-phase extraction provided a simple sample-cleanup procedure before chromatographic analysis.
3
Chromatographic separation used a C8 column with heptafluorobutyric acid as the ion-pair reagent, while detection employed an electrospray source and ion-trap mass analyzer.
4
The method achieved a detection limit of 0.03 mg/kg and a quantitation limit of 0.08 mg/kg, both below European Union maximum residue levels.
5
The method showed good reproducibility, with run-to-run precision of 8–9% and day-to-day precision of 9–13%, and was demonstrated for pears, grapes, purées, and fruit juices.

Chlormequat residues in fruit samples, including pears, grapes, purées, and pear, apple, and grape juices

Determination performance, including detection and quantitation limits, precision, and applicability of chlormequat residue analysis

Publication Details
Publication Date
2001-11-01
Journal
Publisher
ISSN
Access Type
Author Information
Authors
R. Castro
Encarnación Moyano
Maria Teresa Galceran
Explore further