Wireless, Flexible, Ion-Selective Electrode System for Selective and Repeatable Detection of Sodium
Беспроводная гибкая система ион-селективных электродов для селективного и повторяемого обнаружения натрия
2020-06-10
SCID: 54.1/4eeepn8p
Discuss with AI
carbon black–elastomer ion-to-electron transducersaliva sodium detectionselectivity coefficient (Na+ vs K+, -3.0)sodium ion-selective electrode (Na+ ISE)wireless flexible ion-selective electrodes
Figures from the paper
Abstract (AI)
Wireless, flexible, ion-selective electrodes (ISEs) are of great interest in the development of wearable health monitors and clinical systems. Existing film-based electrochemical sensors, however, still have practical limitations due to poor electrical contact and material-interfacial leakage. Here, we introduce a wireless, flexible film-based system with a highly selective, stable, and reliable sodium sensor. A flexible and hydrophobic composite with carbon black and soft elastomer serves as an ion-to-electron transducer offering cost efficiency, design simplicity, and long-term stability. The sensor package demonstrates repeatable analysis of selective sodium detection in saliva with good sensitivity (56.1 mV/decade), stability (0.53 mV/h), and selectivity coefficient of sodium against potassium (-3.0). The film ISEs have an additional membrane coating that provides reinforced stability for the sensor upon mechanical bending. Collectively, the comprehensive study of materials, surface chemistry, and sensor design in this work shows the potential of the wireless flexible sensor system for low-profile wearable applications.
Key Findings
1
A flexible hydrophobic composite of carbon black and soft elastomer functions as an ion-to-electron transducer, providing cost efficiency, design simplicity, and long-term stability.
2
A wireless, flexible film-based ion-selective electrode system for sodium detection was developed to address poor electrical contact and interfacial leakage in existing sensors.
3
An additional membrane coating on the film ISEs reinforces sensor stability under mechanical bending, supporting wearable applications.
4
The sensor achieves repeatable selective sodium detection in saliva with sensitivity of 56.1 mV/decade.
5
The sensor demonstrates stability of 0.53 mV/h and a sodium-versus-potassium selectivity coefficient of -3.0.
Research Object
Wireless, flexible film-based ion-selective electrode (ISE) system for sodium sensing
Research Subject
Selective, repeatable and stable detection performance of sodium including sensitivity, stability, selectivity against potassium, and mechanical robustness (bending) of the sensor system
Publication Details
Publication Date
2020-06-10
Journal
Publisher
ISSN
Open access PDF
Access Type
Author Information
Download PDF
Subscribe to digest