A feasibility study using radiochromic films for fast neutron 2D passive dosimetry

Исследование возможности применения радиохромных плёнок для быстрой двумерной пассивной дозиметрии быстрых нейтронов
Samuel L. Brady, Rathnayaka Gunasingha, Terry T. Yoshizumi, C. R. Howell, A. S. Crowell, B. Fallin, A. P. Tonchev, Mark W. Dewhirst
2010-08-06

2D passive dosimetryMonte Carlo benchmarkingfast neutron dosimetrymouse phantomradiochromic films
The objective of this paper is threefold: (1) to establish sensitivity of XRQA and EBT radiochromic films to fast neutron exposure; (2) to develop a film response to radiation dose calibration curve and (3) to investigate a two-dimensional (2D) film dosimetry technique for use in establishing an experimental setup for a radiobiological irradiation of mice and to assess the dose to the mice in this setup. The films were exposed to a 10 MeV neutron beam via the (2)H(d,n)(3)He reaction. The XRQA film response was a factor of 1.39 greater than EBT film response to the 10 MeV neutron beam when exposed to a neutron dose of 165 cGy. A film response-to-soft tissue dose calibration function was established over a range of 0-10 Gy and had a goodness of fit of 0.9926 with the calibration data. The 2D film dosimetry technique estimated the neutron dose to the mice by measuring the dose using a mouse phantom and by placing a piece of film on the exterior of the experimental mouse setup. The film results were benchmarked using Monte Carlo and aluminum (Al) foil activation measurements. The radiochromic film, Monte Carlo and Al foil dose measurements were strongly correlated, and the film within the mouse phantom agreed to better than 7% of the externally mounted films. These results demonstrated the potential application of radiochromic films for passive 2D neutron dosimetry.
1
A film-response-to-soft-tissue-dose calibration curve was established over 0–10 Gy, achieving a goodness of fit of 0.9926.
2
A passive two-dimensional radiochromic-film dosimetry technique was developed to estimate neutron doses in a mouse irradiation setup.
3
Dose measured inside a mouse phantom agreed with externally mounted-film measurements to within 7%, demonstrating potential for fast neutron 2D passive dosimetry.
4
Film, Monte Carlo, and aluminum-foil activation measurements were strongly correlated, validating the film-based dose estimates.
5
XRQA radiochromic film showed 1.39-fold greater response than EBT film after exposure to a 10 MeV neutron beam delivering 165 cGy.

radiochromic XRQA and EBT films exposed to a 10 MeV fast-neutron beam, including a mouse-irradiation setup and phantom

sensitivity, dose-response calibration, and accuracy of passive two-dimensional neutron dosimetry for estimating dose to mice

Publication Details
Publication Date
2010-08-06
Journal
Publisher
ISSN
Access Type
Author Information
Authors
Samuel L. Brady
Rathnayaka Gunasingha
Terry T. Yoshizumi
C. R. Howell
A. S. Crowell
B. Fallin
A. P. Tonchev
Mark W. Dewhirst
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%