A membrane’s blueprint: In silico investigation of fluid flow and molecular transport as a function of membrane design parameters in organ-on-a-chip

Конструкция мембраны: численное исследование течения жидкости и молекулярного транспорта в зависимости от параметров конструкции мембраны в системе «орган-на-чипе»
Pedro Duarte Menezes, Nikolaj Gadegaard, S. Hecht, A. Hunter
2023-12-21

finite element modellingmicrofluidic flowmolecular transportorgan-on-a-chipporous membranes
Following the rapid growth of Organ-on-a-Chip (OoaC) technology, porous membranes have become essential components for in vitro tissue barrier models. Nonetheless, literature highlights lacking knowledge on their integration and effect on microfluidic devices. Therefore, we conducted finite element modelling (FEM) to characterize the influence of membrane, channel geometry, flow and diffusion parameters, in modelling flow rate, shear stress, transient transport and steady state molecular concentration. This analysis was performed for four different conditions based on single channel (SCP) and parallel perfusion (PP). It was found that membrane and geometry parameters are crucial in determining flow and shear for SCP. However, for PP, flow and shear are predominantly governed by the inlet flow rate. Although the transient behaviour is well-controlled within SCP and PP, only PP allows modelling the steady state concentration distribution. It is highlighted that: (1) the pore radius has great influence on flow and shear; (2) a shallow cell channel and a long membrane are capable of establishing different levels of shear on opposing surfaces of the same channel; (3) the membrane thickness, membrane length, height of the cell and flow channels, and inlet flow rate provide good control over transient transport; (4) the membrane length and inlet flow rate enable changing the concentration from a uniform distribution to a complete heterogeneous state across the device. Experimental assays were performed to support the FEM and evidence its significance for OoaC applications. Ultimately, extensive, and systematic guidelines are provided on designing future OoaC devices with integrated porous membranes.
1
Both single-channel and parallel-perfusion configurations control transient transport, but only parallel perfusion enables modelling steady-state concentration distributions.
2
Experimental assays supported the FEM results, yielding systematic design guidelines for future organ-on-a-chip devices with integrated porous membranes.
3
Finite element modelling systematically quantified how membrane, channel geometry, flow, and diffusion parameters affect fluid flow, shear stress, and molecular transport in organ-on-a-chip devices.
4
For single-channel perfusion, membrane and geometric parameters strongly determine flow and shear, whereas parallel perfusion is governed predominantly by inlet flow rate.
5
Membrane thickness, membrane length, channel heights, and inlet flow rate control transient transport; membrane length and inlet flow rate tune concentration patterns from uniform to completely heterogeneous.
6
Pore radius strongly influences flow and shear, while a shallow cell channel and long membrane can create different shear levels on opposing channel surfaces.

porous membrane-integrated organ-on-a-chip microfluidic devices with single-channel and parallel-perfusion configurations

the effects of membrane design, channel geometry, flow, and diffusion parameters on fluid flow, shear stress, transient molecular transport, and steady-state concentration distribution

Publication Details
Publication Date
2023-12-21
Journal
Publisher
ISSN
Cited by
5
Access Type
Author Information
Authors
Pedro Duarte Menezes
Nikolaj Gadegaard
S. Hecht
A. Hunter
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%