Real-time monitoring of luminescent lifetime changes of PtOEP oxygen sensing film with LED/photodiode-based time-domain lifetime device
Мониторинг в реальном времени изменений времени жизни люминесценции кислородочувствительной пленки PtOEP с помощью приборa для временной регистрации времени жизни на базе светодиода/фотодиода
2009-01-01
SCID: 54.1/4zz25vgx
Discuss with AI
LED/photodiode time-domain luminescent lifetime devicelifetime-based Stern-Volmer analysisoxygen sensing film (PtOEP in IMPEK-C)phosphorescence lifetime monitoring (75 μs to <2 μs)phosphorescent platinum octaethylporphyrin (PtOEP)
Figures from the paper
Abstract (AI)
A cost-effective LED/photodiode(PD)-based time-domain luminescent lifetime measuring device with rugged electronics and simplified algorithms was assembled and successfully used to characterize oxygen sensing films, by continuously monitoring phosphorescence lifetime changes of phosphorescent platinum octaethylporphyrin (PtOEP) in cardo poly(aryl ether ketone) polymer (IMPEK-C) vs. variation of the oxygen partial pressure in a gas mixture (O(2)/N(2)). The results determined by both phosphorescence lifetime and intensity monitoring were compared and the lifetime mode gave results which are in good agreement with the intensity mode. The lifetime-based linear Stern-Volmer plot indicates that the PtOEP molecules are nearly homogeneously distributed in the sensing film. The phosphorescent lifetime of the PtOEP film changes from 75 micros in neat N(2) to less than 2 micros in neat O(2). The sensing system (by combination of the PtOEP sensing film with the home-assembled lifetime device) gives a high lifetime-based O(2) sensing resolution, e.g. about 2 micros Torr(-1) for low O(2) concentration (below 3.5% O(2), V/V). This feasible lifetime device configuration is affordable to most sensor laboratories and the device may facilitate the study of O(2) sensing material with the continuous lifetime monitoring method.
Key Findings
1
A cost-effective LED/photodiode-based time-domain luminescent lifetime device with rugged electronics and simplified algorithms was assembled for monitoring PtOEP oxygen sensing films.
2
Lifetime-based Stern–Volmer plot was linear, indicating near-homogeneous distribution of PtOEP molecules in the sensing film.
3
Phosphorescence lifetime measurements agreed well with intensity-based monitoring for PtOEP in IMPEK-C across varying O2/N2 mixtures.
4
PtOEP phosphorescent lifetime ranged from 75 μs in pure N2 to less than 2 μs in pure O2.
5
The combined PtOEP film and home-assembled lifetime device achieved high lifetime-based O2 sensing resolution, e.g., about 2 μs Torr−1 for O2 concentrations below 3.5% (V/V).
Research Object
Phosphorescent platinum octaethylporphyrin (PtOEP) oxygen-sensing film in IMPEK-C polymer
Research Subject
Real-time changes of phosphorescence lifetime (oxygen-dependent lifetime quenching) of the PtOEP sensing film under varying oxygen partial pressure measured with an LED/photodiode time-domain lifetime device, and its sensing performance (resolution, Stern–Volmer behavior)
Publication Details
Publication Date
2009-01-01
Journal
Publisher
ISSN
Access Type
Author Information
Download PDF
Subscribe to digest