Wrist-Worn Wearables for Monitoring Heart Rate and Energy Expenditure While Sitting or Performing Light-to-Vigorous Physical Activity: Validation Study

Носимые устройства на запястье для мониторинга частоты сердечных сокращений и энергозатрат в положении сидя или при выполнении физических упражнений от низкой до высокой интенсивности: валидационное исследование
Peter Düking, Laura Giessing, Marie Ottilie Frenkel, Karsten Koehler, Hans‐Christer Holmberg, Billy Sperlich
2020-01-24

energy expenditureheart rate monitoringindirect calorimetryvalidation studywrist-worn wearables
BACKGROUND: Physical activity reduces the incidences of noncommunicable diseases, obesity, and mortality, but an inactive lifestyle is becoming increasingly common. Innovative approaches to monitor and promote physical activity are warranted. While individual monitoring of physical activity aids in the design of effective interventions to enhance physical activity, a basic prerequisite is that the monitoring devices exhibit high validity. OBJECTIVE: Our goal was to assess the validity of monitoring heart rate (HR) and energy expenditure (EE) while sitting or performing light-to-vigorous physical activity with 4 popular wrist-worn wearables (Apple Watch Series 4, Polar Vantage V, Garmin Fenix 5, and Fitbit Versa). METHODS: While wearing the 4 different wearables, 25 individuals performed 5 minutes each of sitting, walking, and running at different velocities (ie, 1.1 m/s, 1.9 m/s, 2.7 m/s, 3.6 m/s, and 4.1 m/s), as well as intermittent sprints. HR and EE were compared to common criterion measures: Polar-H7 chest belt for HR and indirect calorimetry for EE. RESULTS: While monitoring HR at different exercise intensities, the standardized typical errors of the estimates were 0.09-0.62, 0.13-0.88, 0.62-1.24, and 0.47-1.94 for the Apple Watch Series 4, Polar Vantage V, Garmin Fenix 5, and Fitbit Versa, respectively. Depending on exercise intensity, the corresponding coefficients of variation were 0.9%-4.3%, 2.2%-6.7%, 2.9%-9.2%, and 4.1%-19.1%, respectively, for the 4 wearables. While monitoring EE at different exercise intensities, the standardized typical errors of the estimates were 0.34-1.84, 0.32-1.33, 0.46-4.86, and 0.41-1.65 for the Apple Watch Series 4, Polar Vantage V, Garmin Fenix 5, and Fitbit Versa, respectively. Depending on exercise intensity, the corresponding coefficients of variation were 13.5%-27.1%, 16.3%-28.0%, 15.9%-34.5%, and 8.0%-32.3%, respectively. CONCLUSIONS: The Apple Watch Series 4 provides the highest validity (ie, smallest error rates) when measuring HR while sitting or performing light-to-vigorous physical activity, followed by the Polar Vantage V, Garmin Fenix 5, and Fitbit Versa, in that order. The Apple Watch Series 4 and Polar Vantage V are suitable for valid HR measurements at the intensities tested, but HR data provided by the Garmin Fenix 5 and Fitbit Versa should be interpreted with caution due to higher error rates at certain intensities. None of the 4 wrist-worn wearables should be employed to monitor EE at the intensities and durations tested.
1
Apple Watch Series 4 showed the highest heart-rate validity, with the smallest standardized errors and coefficients of variation across exercise intensities.
2
Energy-expenditure estimates were considerably less precise than heart-rate measurements across all wearables, with coefficients of variation ranging from 8.0% to 34.5%.
3
For energy expenditure, Polar Vantage V generally showed lower standardized errors than the Apple Watch, whereas Fitbit Versa had the lowest coefficient of variation at some intensities.
4
Heart-rate accuracy varied substantially among devices; Fitbit Versa had the largest errors, with coefficients of variation reaching 19.1%.
5
The study validated four wrist-worn wearables for heart-rate and energy-expenditure monitoring during sitting, walking, running, and intermittent sprints in 25 individuals.

Four popular wrist-worn wearable devices (Apple Watch Series 4, Polar Vantage V, Garmin Fenix 5, and Fitbit Versa) used during sitting and light-to-vigorous physical activity

Validity of heart-rate and energy-expenditure monitoring across different physical-activity intensities

Publication Details
Publication Date
2020-01-24
Journal
Publisher
ISSN
Access Type
Author Information
Authors
Peter Düking
Laura Giessing
Marie Ottilie Frenkel
Karsten Koehler
Hans‐Christer Holmberg
Billy Sperlich
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%