UHMWPE for arthroplasty: past or future?

Сверхвысокомолекулярный полиэтилен в эндопротезировании суставов: прошлое или будущее?
Elena Maria Brach Del Prever, Alessandro Bistolfi, Pierangiola Bracco, Luigi Costa
2008-12-23

UHMWPE arthroplastycrosslinked polyethylenedegradative oxidationvitamin E stabilizationwear debris osteolysis
Wear debris related osteolysis is recognised as being the main cause of failure in joint replacements based on UHMWPE inserts. However, many solutions and "new" polyethylenes have been suggested in order to address this issue. This review discusses "historical" issues associated with UHMWPE, such as oxidation, sterilization method and storage, as well as "new" topics, such as crosslinking and stabilization. The final aim is to aid orthopaedic surgeons in their selection of polyethylene inserts and in the information given to the patients. The main problem for the polymer is degradative oxidation, which is caused by the combination of the irradiation used for sterilization and oxygen, and which leads to a decrease in wear resistance and mechanical properties. Irradiation and packaging in the absence of oxygen can only reduce the oxidation, while sterilization with gas (EtO or gas plasma) is the only method that effectively eliminates it. Manufacturing processes are of great relevance to the clinical duration and must be considered by surgeons. Crosslinked polyethylene has been developed for joint inserts due to its superior wear resistance compared to conventional UHMWPE; to prevent the oxidation, crosslinked polyethylene requires post-irradiation thermal treatment, which reduces its mechanical properties and which depends on the producer. Several good clinical results from the use of crosslinked acetabular cups have reported at mid-term, while early results for knee replacements are also encouraging. Recently, the use of the antioxidant vitamin E (alpha-tocopherol) has been introduced for joint prostheses in order to prevent the oxidation of both crosslinked and noncrosslinked UHMWPE.
1
Crosslinked polyethylene provides superior wear resistance to conventional UHMWPE but requires post-irradiation thermal treatment that can reduce mechanical properties.
2
Degradative oxidation, promoted by irradiation sterilization combined with oxygen exposure, reduces UHMWPE wear resistance and mechanical properties.
3
Mid-term clinical results for crosslinked acetabular cups and early knee-replacement results are encouraging; vitamin E has been introduced to limit oxidation in both crosslinked and noncrosslinked UHMWPE.
4
Oxygen-free irradiation and packaging reduce oxidation, whereas EtO or gas-plasma sterilization effectively eliminates irradiation-related oxidation.
5
Wear-debris-induced osteolysis is identified as the primary failure mechanism in joint replacements using UHMWPE inserts.

UHMWPE polyethylene inserts used in joint replacements

The effects of oxidation, sterilization, storage, crosslinking, stabilization, and manufacturing on the wear resistance, mechanical properties, and clinical performance of UHMWPE inserts

Publication Details
Publication Date
2008-12-23
Journal
Publisher
ISSN
Access Type
Author Information
Authors
Elena Maria Brach Del Prever
Alessandro Bistolfi
Pierangiola Bracco
Luigi Costa
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%