Corrosion equipment for fatigue testing of bioresorbable magnesium alloys

Оборудование для коррозионных испытаний на усталость биоадсорбируемых магниевых сплавов
M. L. Linderov, Alexander Brilevsky, Д. Л. Мерсон
2025-09-25

MA14 magnesium alloybioresorbable magnesium alloyscorrosion-fatigue testing setuppH-controlled corrosion chamberphysiological solution corrosion testing
Currently, there is no generally accepted standard setup for measurements the fatigue limits of biomedical magnesium alloy samples in corrosive environments. Usually every research group uses assemble its own testing system, which does not always take into account many technical factors that significantly affect the final result of the experiment. In this paper, we propose a setup that can be assembled on the basis of a standard fatigue machine for measurements the corrosion-fatigue durability of bioresorbable magnesium alloys in various physiological solutions. It’s main components include: a testing machine with a slot plate required for rigid attachment of the chamber; polyamide grips providing galvanic isolation; a peristaltic pump for pumping the medical solution used; an ionometer connected to a computer to record changes in pH and temperature; a thermostat to maintain the temperature in the chamber within a given range; an additional solution fed into the chamber through a needle using a small peristaltic pump when a given pH level is exceeded, which allows maintaining the pH within the acceptable range of values. The setup uses corset-type samples, although other options are also acceptable, and provides the ability to install a camera to visualize corrosion processes on witness samples. The test system allows achieving good repeatability and reliability of the obtained results, which was experimentally verified and confirmed on magnesium alloy MA14 of different melts, selected as a model. The proposed setup can be used to assess the fatigue life of magnesium alloys for medical purposes. It is not difficult to implement in ordinary factory laboratory conditions.
1
A modular corrosion-fatigue test setup is proposed that can be assembled from a standard fatigue machine for biomedical magnesium alloys in physiological solutions.
2
An auxiliary peristaltic pump injects solution via a needle when pH exceeds a set level, enabling maintenance of pH within an acceptable range during tests.
3
Experimental verification on magnesium alloy MA14 from different melts demonstrated good repeatability and reliability of corrosion-fatigue results.
4
Key components include a slot plate for rigid chamber attachment, polyamide grips for galvanic isolation, a peristaltic pump, an ionometer for pH/temperature logging, and a thermostat for temperature control.
5
The proposed system is easy to implement in standard factory laboratory conditions and can be used to assess fatigue life of medical magnesium alloys.
6
The setup accommodates corset-type samples and allows camera installation to visualize corrosion on witness samples.

Corrosion-fatigue testing setup assembled on a standard fatigue machine for bioresorbable magnesium alloys in physiological solutions

Ability of the setup to measure corrosion-fatigue durability/fatigue life of bioresorbable magnesium alloys under controlled corrosive (physiological) conditions, including repeatability, reliability, pH and temperature control, galvanic isolation, and visualization of corrosion processes

Publication Details
Publication Date
2025-09-25
Journal
Publisher
ISSN
Cited by
0
Access Type
Author Information
Authors
M. L. Linderov
Alexander Brilevsky
Д. Л. Мерсон
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%