An ionic liquid-based sensor for diclofenac determination in water

John Cassidy, Emma Brennan, Pauline Futvoie, Benjamin Schazmann
2017-05-03

This paper details a miniaturised, solid state ion-selective electrode selective for diclofenac. The sensor comprises a novel ionic liquid electroactive material – an imidazolium–diclofenac ion associate. The ion associate is present in a plasticised poly(vinyl chloride) (PVC) membrane on planar carbon electrodes, with an intermediate poly(3,4-ethylenedioxythiophene) layer. The sensitivity and selectivity of the sensor were determined using chronopotentiometric methods. In response to diclofenac, a slope of −53.3 ± 3.6 mV/dec was observed. A limit of detection of 2.90 × 10−3 g L−1 is reported, with a linear range of 3.18 × 10−3 g L−1 to 3.18 g L−1. The sensors show good selectivity towards diclofenac against pertinent interferent molecules, with a response time of <15 s.