Magnetic Levitation for Soft-Tethered Capsule Colonoscopy Actuated With a Single Permanent Magnet: A Dynamic Control Approach

Магнитная левитация мягко привязанной капсулы для колоноскопии, приводимой в действие одним постоянным магнитом: подход к динамическому управлению
Giovanni Pittiglio, Lavinia Barducci, James Martin, Joseph Norton, Carlo Alberto Avizzano, Keith L. Obstein, Pietro Valdastri
2019-01-24

colon phantom navigationmagnetic levitationnonlinear backstepping controlsingle permanent magnetsoft-tethered capsule colonoscopy
The present letter investigates a novel control approach for magnetically driven soft-tethered capsules for colonoscopy-a potentially painless approach for colon inspection. The focus of this work is on a class of devices composed of a magnetic capsule endoscope actuated by a single external permanent magnet. Actuation is achieved by manipulating the external magnet with a serial manipulator, which in turn produces forces and torques on the internal magnetic capsule. We propose a control strategy which, counteracting gravity, achieves levitation of the capsule. This technique, based on a nonlinear backstepping approach, is able to limit contact with the colon walls, reducing friction, avoiding contact with internal folds, and facilitating the inspection of nonplanar cavities. The approach is validated on an experimental setup, which embodies a general scenario faced in colonoscopy. The experiments show that we can attain 19.5% of contact with the colon wall, compared to the almost 100% of previously proposed approaches. Moreover, we show that the control can be used to navigate the capsule through a more realistic environment-a colon phantom-with reasonable completion time.
1
A nonlinear backstepping controller enables gravity-compensated magnetic levitation of a soft-tethered capsule using a single external permanent magnet.
2
Experiments achieved 19.5% colon-wall contact, compared with nearly 100% for previously proposed approaches.
3
Levitation reduces capsule contact with colon walls, thereby lowering friction, avoiding internal folds, and supporting inspection of nonplanar cavities.
4
The control strategy successfully navigated the capsule through a realistic colon phantom within a reasonable completion time.

soft-tethered magnetic capsule endoscope actuated by a single external permanent magnet for colonoscopy

dynamic levitation and navigation control that counteracts gravity and reduces capsule contact with colon walls

Publication Details
Publication Date
2019-01-24
Journal
Publisher
ISSN
Access Type
Author Information
Authors
Giovanni Pittiglio
Lavinia Barducci
James Martin
Joseph Norton
Carlo Alberto Avizzano
Keith L. Obstein
Pietro Valdastri
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%