Evaluation of In-room Volumetric Imaging Doses for Image-guided Radiotherapy: A Multi-institutional Study

Оценка доз объемной визуализации в помещении для направляемой изображением радиотерапии: многоучрежденческое исследование
Yusuke Sakai, Hajime Monzen, Yoshiki Takei, Hiroyuki Kosaka, Kenji Nakamura, Yuya Yanagi, Kazuki Wakabayashi, Makoto Hosono, Yasumasa Nishimura
2023-04-01

diagnostic reference levelsimage-guided radiotherapyimaging dosein-room volumetric imagingpelvic phantom
Aims: We investigated imaging dose and noise under clinical scan conditions at multiple institutions using a simple and unified method, and demonstrated the need for diagnostic reference levels in image-guided radiotherapy (IGRT). Materials and Methods: Nine cone-beam and helical computed tomography (CT) scanners (Varian, Elekta, Accuray Inc., and BrainLAB) from seven institutions were investigated in this study. The weighted cone-beam dose index (CBDI w ) was calculated for head and pelvic protocols using a 100 mm pencil chamber under the conditions used in actual clinical practice at each institution. Cone-beam CT image noise was evaluated using polymethylmethacrylate head and body phantoms with diameters of 16 and 32 cm, respectively. Results: For head and pelvic protocols, CBDI w values ranged from 0.94–6.59 and 1.47–20.9 mGy, respectively. Similarly, standard deviation (SD) values ranged from 9.3–34.0 and 26.9–97.4 HU, respectively. The SD values tended to increase with decreasing imaging dose ( r = −0.33 and −0.61 for the head and pelvic protocols, respectively). Conclusions: Among the nine machines, the imaging dose for high imaging dose institutions was approximately 20 mGy to the pelvic phantom, and there was a 14-fold difference in dose compared with the other institutions. These results suggest the need to establish DRLs for IGRT to guide clinical decision-making.
1
Across nine machines, imaging dose to a pelvic phantom varied widely between institutions.
2
Findings indicate a need to establish diagnostic reference levels (DRLs) for IGRT to guide clinical decision-making.
3
High-imaging-dose institutions delivered approximately 20 mGy to the pelvic phantom.
4
There was a 14-fold difference in imaging dose between high-dose and low-dose institutions.

In-room volumetric imaging for image-guided radiotherapy across multiple institutions (imaging machines and pelvic phantom measurements)

Imaging radiation dose levels delivered by in-room volumetric imaging systems for IGRT, including inter-institutional dose variation and the need to establish diagnostic reference levels (DRLs)

Publication Details
Publication Date
2023-04-01
Journal
Publisher
ISSN
Cited by
1
Access Type
Author Information
Authors
Yusuke Sakai
Hajime Monzen
Yoshiki Takei
Hiroyuki Kosaka
Kenji Nakamura
Yuya Yanagi
Kazuki Wakabayashi
Makoto Hosono
Yasumasa Nishimura
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%