Evaluation of In-room Volumetric Imaging Doses for Image-guided Radiotherapy: A Multi-institutional Study
Оценка доз объемной визуализации в помещении для направляемой изображением радиотерапии: многоучрежденческое исследование
2023-04-01
SCID: 54.1/bj6u3zrs
Discuss with AI
diagnostic reference levelsimage-guided radiotherapyimaging dosein-room volumetric imagingpelvic phantom
Figures from the paper
Abstract (AI)
Aims: We investigated imaging dose and noise under clinical scan conditions at multiple institutions using a simple and unified method, and demonstrated the need for diagnostic reference levels in image-guided radiotherapy (IGRT). Materials and Methods: Nine cone-beam and helical computed tomography (CT) scanners (Varian, Elekta, Accuray Inc., and BrainLAB) from seven institutions were investigated in this study. The weighted cone-beam dose index (CBDI w ) was calculated for head and pelvic protocols using a 100 mm pencil chamber under the conditions used in actual clinical practice at each institution. Cone-beam CT image noise was evaluated using polymethylmethacrylate head and body phantoms with diameters of 16 and 32 cm, respectively. Results: For head and pelvic protocols, CBDI w values ranged from 0.94–6.59 and 1.47–20.9 mGy, respectively. Similarly, standard deviation (SD) values ranged from 9.3–34.0 and 26.9–97.4 HU, respectively. The SD values tended to increase with decreasing imaging dose ( r = −0.33 and −0.61 for the head and pelvic protocols, respectively). Conclusions: Among the nine machines, the imaging dose for high imaging dose institutions was approximately 20 mGy to the pelvic phantom, and there was a 14-fold difference in dose compared with the other institutions. These results suggest the need to establish DRLs for IGRT to guide clinical decision-making.
Key Findings
1
Across nine machines, imaging dose to a pelvic phantom varied widely between institutions.
2
Findings indicate a need to establish diagnostic reference levels (DRLs) for IGRT to guide clinical decision-making.
3
High-imaging-dose institutions delivered approximately 20 mGy to the pelvic phantom.
4
There was a 14-fold difference in imaging dose between high-dose and low-dose institutions.
Research Object
In-room volumetric imaging for image-guided radiotherapy across multiple institutions (imaging machines and pelvic phantom measurements)
Research Subject
Imaging radiation dose levels delivered by in-room volumetric imaging systems for IGRT, including inter-institutional dose variation and the need to establish diagnostic reference levels (DRLs)