Fragmented sleep due to frequent awakenings represents a major cause of impaired daytime performance and adverse health outcomes. Currently, the gold standard for studying and assessing sleep fragmentation is polysomnography (PSG). Here, we propose an alternative method for real-time detection of nocturnal awakening via ballistocardiography using an unobtrusive polyvinylidene fluoride (PVDF) film sensor on a bed mattress. From ballistocardiogram, heart rate and body movement information were extracted to develop an algorithm for classifying sleeping and awakening epochs. In total, ten normal subjects (mean age 38.7 ± 14.6 years) and ten patients with obstructive sleep apnea (OSA) (mean age 44.2 ± 16.5 years) of varying symptom severity participated in this study. Our study detected awakening epochs with an average sensitivity of 85.3% and 85.2%, specificity of 98.4% and 97.7%, accuracy of 97.4% and 96.5%, and Cohen's kappa coefficient of 0.83 and 0.81 for normal subjects and OSA patients, respectively. Also, sleep efficiency was estimated using detected awakening epochs and then compared with PSG results. Mean absolute errors in sleep efficiency were 1.08% and 1.44% for normal subjects and OSA patients, respectively. The results presented here indicate that our suggested method could be reliably applied to real-time nocturnal awakening detection and sleep efficiency estimation. Furthermore, our method may ultimately be an effective tool for long-term, home monitoring of sleep-wake behavior.
Key Findings
1
An unobtrusive polyvinylidene fluoride film sensor embedded in a mattress enabled real-time nocturnal awakening detection using ballistocardiographic heart-rate and movement information.
2
In obstructive sleep apnea patients, awakening detection achieved 85.2% sensitivity, 97.7% specificity, 96.5% accuracy, and Cohen’s kappa of 0.81.
3
Sleep efficiency estimated from detected awakenings differed from polysomnography by mean absolute errors of 1.08% in normal subjects and 1.44% in OSA patients.
4
The findings support reliable real-time awakening detection and sleep-efficiency estimation, with potential for long-term home monitoring of sleep-wake behavior.
5
The method achieved 85.3% sensitivity, 98.4% specificity, 97.4% accuracy, and Cohen’s kappa of 0.83 in normal subjects.
Research Object
Nocturnal sleep–wake behavior in normal subjects and patients with obstructive sleep apnea (OSA), measured via unobtrusive ballistocardiography
Research Subject
Real-time detection of nocturnal awakening and estimation of sleep efficiency from heart-rate and body-movement information extracted from ballistocardiograms
Publication Details
Publication Date
2014-01-01
Journal
Publisher
ISSN
Access Type
Author Information
Download PDF
Subscribe to digest