Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges
Датчики непрерывного мониторинга глюкозы: алгоритмические проблемы прошлого, настоящего и будущего
2016-12-09
SCID: 54.1/df4u73vk
Discuss with AI
artificial pancreascontinuous glucose monitoringglucose predictionhypoglycemia preventionsignal processing
Figures from the paper
Abstract (AI)
Continuous glucose monitoring (CGM) sensors are portable devices that allow measuring and visualizing the glucose concentration in real time almost continuously for several days and are provided with hypo/hyperglycemic alerts and glucose trend information. CGM sensors have revolutionized Type 1 diabetes (T1D) management, improving glucose control when used adjunctively to self-monitoring blood glucose systems. Furthermore, CGM devices have stimulated the development of applications that were impossible to create without a continuous-time glucose signal, e.g., real-time predictive alerts of hypo/hyperglycemic episodes based on the prediction of future glucose concentration, automatic basal insulin attenuation methods for hypoglycemia prevention, and the artificial pancreas. However, CGM sensors' lack of accuracy and reliability limited their usability in the clinical practice, calling upon the academic community for the development of suitable signal processing methods to improve CGM performance. The aim of this paper is to review the past and present algorithmic challenges of CGM sensors, to show how they have been tackled by our research group, and to identify the possible future ones.
Key Findings
1
CGM has improved type 1 diabetes management when used alongside self-monitoring blood glucose systems.
2
Continuous glucose data enabled predictive hypoglycemia and hyperglycemia alerts, automatic basal-insulin attenuation, and artificial-pancreas development.
3
Continuous glucose monitoring sensors provide near-continuous real-time glucose measurements, trend information, and hypo-/hyperglycemia alerts over several days.
4
Limited CGM accuracy and reliability have hindered clinical adoption, motivating signal-processing methods to improve sensor performance.
5
The paper reviews past and current CGM algorithmic challenges, summarizes the research group's solutions, and identifies future challenges.
Research Object
Continuous glucose monitoring (CGM) sensors and their glucose-monitoring data streams
Research Subject
Algorithmic challenges in improving CGM accuracy, reliability, signal processing, and glucose prediction for diabetes management
Publication Details
Publication Date
2016-12-09
Journal
Sensors
Publisher
Multidisciplinary Digital Publishing Institute
ISSN
1424-8220
Cited by
192
Open access PDF
Access Type
Author Information
Download PDF
Subscribe to digest
References available in scid.ai3
In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes2009
Circadian Variability of Insulin Sensitivity: Physiological Input for In Silico Artificial Pancreas2014
Hypoglycemia Prevention via Pump Attenuation and Red-Yellow-Green “Traffic” Lights Using Continuous Glucose Monitoring and Insulin Pump Data2010