Influence of head orientation and implantation site of a novel transcutaneous bone conduction implant on MRI metal artifact reduction sequence

Влияние ориентации головы и места имплантации нового чрескожного имплантата костной проводимости на последовательность МРТ для уменьшения металлических артефактов
Lukas Anschuetz, Wilhelm Wimmer, Arsany Hakim, Claus Kiefer, Manuela Pastore‐Wapp, Marco Caversaccio, Franca Wagner, Georgios Mantokoudis, Emile Talon
2022-01-24

MRI metal artifact reductionSEMAC-VAT WARPbone conduction implanthead orientationimplantation site
PURPOSE: The use of magnetic resonance imaging (MRI) is often limited in patients with auditory implants because of the presence of metallic components and magnets. The aim of this study was to evaluate the clinical usefulness of a customized MRI sequence for metal artifact suppression in patients with BONEBRIDGETM BCI 602 implants (MED-EL, Innsbruck, Austria), the successor of the BCI 601 model. METHODS: Using our in-house developed and customized metal artifact reduction sequence (SEMAC-VAT WARP), MRI artifacts were evaluated qualitatively and quantitatively. MRI sequences were performed with and without artifact reduction on two whole head specimens with and without the BCI 602 implant. In addition, the influence of two different implantation sites (mastoid versus retrosigmoid) and head orientation on artifact presence was investigated. RESULTS: Artifact volume was reduced by more than the 50%. Results were comparable with those obtained with the BCI 601, showing no significant differences in the dimensions of artifacts caused by the implant. CONCLUSION: SEMAC-VAT WARP was once more proved to be efficient at reducing metal artifacts on MR images. The dimensions of artifacts associated with the BCI 602 are not smaller than those caused by the BCI 601.
1
Artifact dimensions caused by the BCI 602 were not significantly different from those caused by the predecessor BCI 601.
2
Artifact reduction was demonstrated in whole-head specimens with and without BCI 602 implantation, supporting the sequence’s clinical usefulness.
3
SEMAC-VAT WARP was confirmed to efficiently suppress MRI metal artifacts associated with the BCI 602 implant.
4
The customized SEMAC-VAT WARP MRI sequence reduced metal artifact volume by more than 50% in specimens with BONEBRIDGE BCI 602 implants.
5
The study investigated how mastoid versus retrosigmoid implantation sites and head orientation influence MRI artifact presence.

BONEBRIDGE BCI 602 transcutaneous bone conduction implants in whole head specimens, with mastoid or retrosigmoid implantation sites

MRI metal artifacts associated with BCI 602 implants and their reduction as a function of customized SEMAC-VAT WARP sequence, implantation site, and head orientation

Publication Details
Publication Date
2022-01-24
Journal
Publisher
ISSN
Cited by
6
Access Type
Author Information
Authors
Lukas Anschuetz
Wilhelm Wimmer
Arsany Hakim
Claus Kiefer
Manuela Pastore‐Wapp
Marco Caversaccio
Franca Wagner
Georgios Mantokoudis
Emile Talon
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%