Two new separation schemes for the group isolation of rare earth elements (REE) from biological and other matrices and their determination by ICP-MS, NAA and chromatographic methods

Две новые схемы разделения для группового выделения редкоземельных элементов (РЗЭ) из биологических и других матриц и их определения методами ИСП-МС, НАА и хроматографии
R. Dybczyński, Zbigniew Samczyński, Iwona Bartosiewicz, Krzysztof Kulisa, Halina Polkowska–Motrenko, Marta Pyszynska, Iga Zuba
2017-07-22

certified reference materialsinductively coupled plasma mass spectrometryinstrumental neutron activation analysision exchange chromatographyrare earth elements
Abstract Two new group separation schemes, based on ion exchange chromatography, for the selective and quantitative isolation of rare earth elements (REE) from accompanying elements, were devised. After checking their performance with the aid of radioactive tracers, the schemes were further used together with ICP-MS, NAA and ion exchange chromatography for the determination of Sc, Y, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb and Lu in two certified reference materials (CRMs). The results were compared with another series of analyses, where the REEs were determined directly, i.e. without pre-separation, by instrumental neutron activation analysis (INAA) and inductively coupled plasma mass spectrometry (ICP-MS). It was demonstrated that while direct INAA and ICP-MS in most instances provide reliable results for the majority of REEs, for some elements, notably Sc, Yb and Tm in the cases of ICP-MS and INAA, respectively, systematic errors occur or may potentially occur.
1
Direct INAA and ICP-MS generally produced reliable results for most rare earth elements, but systematic errors occurred or could occur for specific elements.
2
Sc in ICP-MS measurements and Yb and Tm in INAA measurements were particularly susceptible to systematic errors, demonstrating the value of pre-separation.
3
Sc, Y, La, Ce, Pr, Nd, Sm, Eu, Gd, Tb, Dy, Ho, Er, Tm, Yb, and Lu were determined in two certified reference materials.
4
The separation schemes were validated using radioactive tracers and coupled with ICP-MS, neutron activation analysis, and chromatographic determination.
5
Two ion-exchange chromatography schemes were developed for selective and quantitative group isolation of rare earth elements from accompanying matrix elements.

Rare earth elements (REE) in biological and other matrices, including certified reference materials

Selective and quantitative group isolation and determination of REEs, including the accuracy and systematic errors of direct versus pre-separated ICP-MS and INAA measurements

Publication Details
Publication Date
2017-07-22
Journal
Publisher
ISSN
Access Type
Author Information
Authors
R. Dybczyński
Zbigniew Samczyński
Iwona Bartosiewicz
Krzysztof Kulisa
Halina Polkowska–Motrenko
Marta Pyszynska
Iga Zuba
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%