Sensitivity and False Alarm Rate of a Fall Sensor in Long-Term Fall Detection in the Elderly

Чувствительность и частота ложных тревог датчика падения при долгосрочном выявлении падений у пожилых людей
Maarit Kangas, Raija Korpelainen, Irene Vikman, Lars Nyberg, Timo Jämsä
2014-08-13

accelerometry-based sensorfall detectionfalse alarm rateolder adultssensitivity
BACKGROUND: About a third of home-dwelling older people fall each year, and institutionalized older people even report a two- or threefold higher rate for falling. Automatic fall detection systems have been developed to support the independent and secure living of the elderly. Even though good fall detection sensitivity and specificity in laboratory settings have been reported, knowledge about the sensitivity and specificity of these systems in real-life conditions is still lacking. OBJECTIVE: The aim of this study was to evaluate the long-term fall detection sensitivity and false alarm rate of a fall detection prototype in real-life use. METHODS: A total of 15,500 h of real-life data from 16 older people, including both fallers and nonfallers, were monitored using an accelerometry-based sensor system with an implemented fall detection algorithm. RESULTS: The fall detection system detected 12 out of 15 real-life falls, having a sensitivity of 80.0%, with a false alarm rate of 0.049 alarms per usage hour with the implemented real-time system. With minor modification of data analysis the false alarm rate was reduced to 0.025 false alarms per hour, equating to 1 false fall alarm per 40 usage hours. CONCLUSION: These data suggest that automatic accelerometric fall detection systems might offer a tool for improving safety among older people.
1
Minor data-analysis modifications reduced the false alarm rate to 0.025 per hour, equivalent to one false alarm per 40 usage hours.
2
Real-life performance data suggest automatic accelerometric fall detection may improve safety for older people, although laboratory performance does not fully establish field effectiveness.
3
The implemented real-time system produced 0.049 false alarms per usage hour.
4
The study evaluated an accelerometry-based fall detection prototype during 15,500 hours of real-life monitoring in 16 older people.
5
The system detected 12 of 15 real-life falls, achieving 80.0% sensitivity.

An accelerometry-based automatic fall detection system used by older people in real-life conditions

Long-term fall detection sensitivity and false alarm rate

Publication Details
Publication Date
2014-08-13
Journal
Publisher
ISSN
Access Type
Author Information
Authors
Maarit Kangas
Raija Korpelainen
Irene Vikman
Lars Nyberg
Timo Jämsä
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%