Conductive recycled PETg additive manufacturing filament for sterilisable electroanalytical healthcare sensors
Проводящий филамент из переработанного PETg для аддитивного производства стерилизуемых электроаналитических сенсоров для здравоохранения
2024-06-13
SCID: 54.1/p6bqcw6t
Discuss with AI
conductive recycled PETg filamentdifferential pulse voltammetryelectrochemical healthcare sensorsfused filament fabricationuric acid and nitrite detection
Figures from the paper
Abstract (AI)
• Conductive recycled PETg filament for fused filament fabrication is reported. • An optimised activation procedure for electrodes printed from conductive PETg. • Electrodes can be sterilised and therefore applied within a healthcare setting. • Determination of uric acid and nitrite within synthetic urine samples. • Electrodes could be re-used up to 10 times before a significant loss in sensitivity. Current reports of healthcare sensors within literature that use additive manufacturing electrochemistry all utilise conductive PLA, which is unsuitable for widespread use within the industry. Poly(ethylene terephthalate glycol (PETg) is a polymeric material with proven attributes for additive manufacturing due to its thermal and mechanical properties. Likewise, its excellent chemical stability transforms PETg into a desirable alternative for developing healthcare sensing devices. In this work, we report the production, physicochemical and electrochemical characterisations, as well as the electroanalytical performance of an enhanced electrically conductive additive manufacturing filament made with recycled poly(ethylene terephthalate glycol (rPETg) and a combination of carbon black, multi-walled carbon nanotubes and graphene nanoplatelets as conductive fillers. The post-print activation of additive manufactured electrodes from this material is optimised and shown to produce enhanced electrochemical performance compared to non-activated electrodes, with a k 0 of 1.03×10 −3 cm s −1 . The sterilisation for the real application of sensors in the biomedical field is a critical point, the electrodes were submitted to standard UV light treatment showing to be reliable compared to PLA in the determination of uric acid (30–500 µM) and sodium nitrite (0.1–5 mM) within synthetic urine using differential pulse voltammetry and chronoamperometry techniques. A sensitivity and LOD for uric acid of 25.7 µA µM −1 and 0.27 µM, and 52.6 µA mM −1 and 2.69 µM for nitrite were obtained within synthetic urine, respectively. The re-useability of the electrodes was also tested for the detection of uric acid, showing that the electrode could be used up to 10 times before a significant decrease in the results was observed. We demonstrate that a new conductive rPETg with superior electrochemical performance has a prominent place within the development of additive manufactured-printed healthcare sensors due to its ability to be sterilised and re-used, low solution ingress, and its potential to tackle rising costs and plastic waste problems within the healthcare sector.
Key Findings
1
A conductive recycled PETg filament was developed for fused filament fabrication using carbon black, multi-walled carbon nanotubes, and graphene nanoplatelets.
2
Differential pulse voltammetry and chronoamperometry enabled uric acid determination from 30–500 µM and nitrite determination from 0.1–5 mM in synthetic urine.
3
Optimized post-print activation substantially enhanced electrode electrochemical performance, achieving a heterogeneous electron-transfer rate constant (k0) of 1.03×10−3 cm s−1.
4
The electrodes could be reused up to 10 times before experiencing a significant loss in sensitivity.
5
The printed PETg electrodes tolerated standard UV sterilization and performed reliably for healthcare-oriented electroanalysis, offering advantages over conductive PLA.
6
Uric acid measurements achieved a sensitivity of 25.7 µA µM−1 and a detection limit of 0.27 µM, while nitrite measurements achieved 52.6 µA mM−1 and a 2.69 µM detection limit.
Research Object
Sterilisable electrodes and healthcare sensors fabricated from conductive recycled PETg additive-manufacturing filament
Research Subject
Their electrochemical performance, activation, sterilisation compatibility, and reusability for determining uric acid and nitrite in synthetic urine