An investigation of kV CBCT image quality and dose reduction for volume‐of‐interest imaging using dynamic collimation

Исследование качества изображений и снижения дозы при киловольтной конусно-лучевой КТ области интереса с использованием динамической коллимации
James L. Robar, David D. M. Parsons
2015-08-14

contrast-to-noise ratiodynamic collimationiris aperturescatter-to-primary ratiovolume-of-interest CBCT
PURPOSE: The focus of this work was to investigate the improvements in image quality and dose reduction for volume-of-interest (VOI) kilovoltage-cone beam CT (CBCT) using dynamic collimation. METHODS: A prototype iris aperture was used to track a VOI during a CBCT acquisition. The current aperture design is capable of 1D translation as a function of gantry angle and dynamic adjustment of the iris radius. The aperture occupies the location of the bow-tie filter on a Varian On-Board Imager system. CBCT and planar image quality were investigated as a function of aperture radius, while maintaining the same dose to the VOI, for a 20 cm diameter cylindrical water phantom with a 9 mm diameter bone insert centered on isocenter. Corresponding scatter-to-primary ratios (SPR) were determined at the detector plane with Monte Carlo simulation using EGSnrc. Dose distributions for various sizes VOI were modeled using a dynamic BEAMnrc library and DOSXYZnrc. The resulting VOI dose distributions were compared to full-field distributions. RESULTS: SPR was reduced by a factor of 8.4 when decreasing iris diameter from 21.2 to 2.4 cm (at isocenter). Depending upon VOI location and size, dose was reduced to 16%-90% of the full-field value along the central axis plane and down to 4% along the axis of rotation, while maintaining the same dose to the VOI compared to full-field techniques. When maintaining constant dose to the VOI, this change in iris diameter corresponds to a factor increase of approximately 1.6 in image contrast and a factor decrease in image noise of approximately 1.2. This results in a measured gain in contrast-to-noise ratio by a factor of approximately 2.0. CONCLUSIONS: The presented VOI technique offers improved image quality for image-guided radiotherapy while sparing the surrounding volume of unnecessary dose compared to full-field techniques.
1
At equal dose to the VOI, surrounding-tissue dose decreased to 16%–90% of full-field values in the central-axis plane and to 4% along the rotation axis.
2
Dynamic VOI collimation produced an approximately 2.0-fold gain in measured contrast-to-noise ratio compared with full-field imaging.
3
Dynamic iris collimation for VOI kV-CBCT reduced the detector-plane scatter-to-primary ratio by a factor of 8.4 when iris diameter decreased from 21.2 to 2.4 cm.
4
Reducing iris diameter while maintaining VOI dose increased image contrast approximately 1.6-fold and reduced image noise approximately 1.2-fold.
5
The technique improves image quality for image-guided radiotherapy while sparing surrounding tissues from unnecessary radiation dose.

VOI kilovoltage cone-beam CT imaging using dynamic collimation

Image-quality improvement and dose reduction, including changes in scatter-to-primary ratio, contrast, noise, contrast-to-noise ratio, and dose distributions relative to full-field imaging

Publication Details
Publication Date
2015-08-14
Journal
Publisher
ISSN
Cited by
20
Access Type
Author Information
Authors
James L. Robar
David D. M. Parsons
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%