Periodic detection and disinfection maintenance of dental unit waterlines in dental simulation head model laboratories

Периодический контроль и дезинфекционное обслуживание водопроводных линий стоматологических установок в лабораториях с моделями головы для стоматологической симуляции
Boyang Li, Yong Hu, Yiqi Wang, Chengqi Zhang, Zimo Wang, Xujie Peng, Feng Jianying
2025-02-12

chlorine dioxide disinfectiondental simulation laboratoriesdental unit waterlinesmicrobial diversitywater contamination
Dental simulation head model laboratories are crucial for clinical simulation training for stomatological students, yet the maintenance of their dental unit waterlines (DUWLs) has been overlooked. This study investigated water contamination in DUWLs within these laboratories and proposed solutions. Water samples were collected from 12 dental chairs in three laboratories at three time points: the beginning, middle, and end of the semester. At the start of the semester, severe contamination was observed, with colony counts of 11,586[Formula: see text]1715 CFU/ml for high-speed handpieces and 5375[Formula: see text]874 CFU/ml for three ways syringes. As the semester progressed, colony counts gradually decreased but remained above clinical thresholds. Both 20 mg/L organochlorine disinfectant and 20 mg/L chlorine dioxide were effective in reducing bacterial contamination below standard ranges three days post-disinfection. Microbial diversity analysis revealed Proteobacteria and Bacteroidota as the dominant bacterial phyla, with Ascomycota as the dominant fungal phylum. Potentially pathogenic bacteria such as Pseudomonas, Burkholderia-Caballeronia-Paraburkholderia, Ralstonia, Mycobacterium, Legionella, Paenibacillus, Streptomyces, Acinetobacter, and Prevotella, as well as fungi like Fusarium and Penicillium, were detected. Therefore, urgent attention is needed to address DUWL contamination in dental laboratories, and it is recommended to disinfect DUWLs using either 20 mg/L organochlorine disinfectant or 20 mg/L chlorine dioxide every three days.
1
Bacterial colony counts decreased during the semester but remained above clinical thresholds, indicating persistent DUWL contamination without adequate maintenance.
2
Both 20 mg/L organochlorine disinfectant and 20 mg/L chlorine dioxide reduced bacterial contamination below standard ranges three days after disinfection.
3
Dental unit waterlines in simulation head model laboratories showed severe initial contamination, reaching 11,586 ± 1,715 CFU/ml in high-speed handpieces and 5,375 ± 874 CFU/ml in three-way syringes.
4
Potentially pathogenic bacteria and fungi were detected, including Pseudomonas, Mycobacterium, Legionella, Acinetobacter, Fusarium, and Penicillium; disinfection every three days was recommended.
5
Proteobacteria and Bacteroidota dominated bacterial communities, while Ascomycota dominated fungal communities in the contaminated waterlines.

Dental unit waterlines (DUWLs) in dental simulation head model laboratories

Microbial contamination, community composition, and disinfection effectiveness in DUWLs over the semester

Publication Details
Publication Date
2025-02-12
Journal
Publisher
ISSN
Cited by
7
Access Type
Author Information
Authors
Boyang Li
Yong Hu
Yiqi Wang
Chengqi Zhang
Zimo Wang
Xujie Peng
Feng Jianying
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%