Statistical Image Reconstruction Methods for Transmission Tomography
Статистические методы реконструкции изображений для просвечивающей томографии
2010-03-16
SCID: 54.1/qbs6jthz
Discuss with AI
Poisson measurement modelingattenuation mapslow-count transmission scansstatistical image reconstructiontransmission tomography
Figures from the paper
Abstract (AI)
The problem of forming cross-sectional or tomographic images of the attenuation characteristics of objects arises in a variety of contexts, including medical x-ray computed tomography (CT) and nondestructive evaluation of objects in industrial inspection. In the context of emission imaging, such as positron emission tomography (PET) [1, 2], single photon emission computed tomography (SPECT) [3], and related methods used in the assay of containers of radioactive waste [4], it is useful to be able to form "attenuation maps," tomographic images of attenuation coefficients, from which one can compute attenuation correction factors for use in emission image reconstruction. One can measure the attenuating characteristics of an object by transmitting a collection of photons through the object along various paths or "rays" and observing the fraction that pass unabsorbed. From measurements collected over a large set of rays, one can reconstruct tomographic images of the object. Such image reconstruction is the subject of this chapter. In all the above applications, the number of photons one can measure in a transmission scan is limited. In medical x-ray CT, source strength, patient motion, and absorbed dose considerations limit the total x-ray exposure. Implanted objects such as pacemakers also significantly reduce transmissivity and cause severe artifacts [5]. In industrial applications, source strength limitations, combined with the very large attenuation coefficients of metallic objects, often result in a small fraction of photons passing to the detector unabsorbed. In PET and SPECT imaging, the transmission scan only determines a "nuisance" parameter of secondary interest relative to the object's emission properties, so one would like to minimize the transmission scan duration. All the above considerations lead to "low-count" transmission scans. This chapter discusses algorithms for reconstructing attenuation images from low-count transmission scans. In this context, we define low-count to mean that the mean number of photons per ray is small enough that traditional filtered-backproject on (FBP) images, or even methods based on the Gaussian approximation to the distribution of the Poisson measurements (or logarithm thereof), are inadequate. We focus the presentation in the context of PET and SPECT transmission scans, but the methods are generally applicable to all low-count transmission studies. See [6] for an excellent survey of statistical approaches for the emission reconstruction problem.
Key Findings
1
Low photon counts arise from practical limits (x-ray dose, source strength, patient motion, implants, high attenuation objects) motivating specialized algorithms for robust reconstruction.
2
Low-count transmission scans (mean photons per ray small) make traditional filtered-backprojection and Gaussian-approximation methods inadequate for attenuation image reconstruction.
3
Methods focus on PET and SPECT transmission scans but are generally applicable to all low-count transmission tomography problems.
4
Statistical image reconstruction algorithms are developed and discussed specifically to reconstruct attenuation images from low-count transmission data.
Research Object
Attenuation images (tomographic attenuation maps) reconstructed from low-count transmission measurements in transmission tomography
Research Subject
Statistical image reconstruction algorithms and their performance for recovering attenuation coefficients from low-count (Poisson) transmission scans, including limitations of FBP and Gaussian approximations