Duplex Ultrasound Criteria for the Identification of Carotid Stenosis Should Be Laboratory Specific

Критерии дуплексного ультразвукового исследования для выявления стеноза сонных артерий должны быть специфичными для каждой лаборатории
Joseph F. Polak, Anthony D. Whittemore, Karen M. Kuntz, Kenneth Craig Kent, John Joakim Skillman
1997-03-01

carotid stenosisduplex ultrasoundinternal carotid arterylaboratory-specific criteriareceiver operating characteristic curves
BACKGROUND AND PURPOSE: Published criteria for the determination of carotid stenosis have been widely applied by vascular laboratories. We compared two vascular laboratories and their duplex ultrasound (DU) machines in terms of their overall diagnostic performance and the optimal criteria to identify patients who have a 70% to 99% stenosis of the internal carotid artery. METHODS: Measurements of stenosis by DU and angiography were compared for 123 carotid arteries (60 arteries, laboratory A; 63 arteries, laboratory B). Receiver operating characteristic (ROC) curves were created, and the areas under the ROC curves and the optimal criteria for determining a 70% to 99% stenosis were compared. Multiple regression analysis was used to measure the effect of laboratory on the relationship between angiographic stenosis and DU velocity parameters. RESULTS: Areas under the ROC curves were similar for both laboratories (0.89 to 0.90, laboratory A; 0.90 to 0.92, laboratory B). However, the optimal criterion for the identification of a 70% to 99% carotid stenosis was different for each laboratory. For most velocity parameters, based on regression analyses, the predicted percent angiographic stenosis for laboratory A was significantly greater than that for laboratory B. In addition, performance differed between the laboratories when established criteria from the literature were applied. CONCLUSIONS: Two vascular laboratories with similar diagnostic accuracy by ROC analysis have markedly different "optimal" DU criteria. For a given angiographic stenosis, velocities in one laboratory were consistently greater than those in the other laboratory. Laboratory-specific criteria rather than published criteria should be used to identify patients with internal carotid artery stenoses.
1
Applying published duplex ultrasound criteria produced different diagnostic performance across the two laboratories.
2
Despite similar ROC performance, the optimal duplex ultrasound velocity criteria for identifying 70%–99% stenosis differed between the laboratories.
3
Duplex ultrasound criteria for severe internal carotid stenosis should be validated and calibrated specifically for each vascular laboratory and equipment setting.
4
For most velocity parameters, laboratory A measured significantly higher velocities than laboratory B for the same angiographically determined stenosis.
5
Two vascular laboratories showed similar overall diagnostic accuracy for detecting 70%–99% internal carotid stenosis, with ROC areas of 0.89–0.90 and 0.90–0.92.

Duplex ultrasound assessment of internal carotid artery stenosis in two vascular laboratories

Laboratory-specific diagnostic criteria and velocity-parameter relationships for identifying 70%–99% internal carotid artery stenosis

Publication Details
Publication Date
1997-03-01
Journal
Publisher
ISSN
Cited by
90
Access Type
Author Information
Authors
Joseph F. Polak
Anthony D. Whittemore
Karen M. Kuntz
Kenneth Craig Kent
John Joakim Skillman
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%