Photophysical and Computational Exploration of Human Serum Albumin Mediated Sensing of Bilirubin via a D‐π‐A Probe
Фот физическое и вычислительное исследование детектирования билирубина с помощью зонда D–π–A, опосредованного сывороточным альбумином человека
2026-07-01
SCID: 54.1/t8nn3cnp
Discuss with AI
D-π-A probeMMGBSA calculationsbilirubin sensinghuman serum albuminmolecular docking
Figures from the paper
Abstract (AI)
Background : Bilirubin is an essential biomarker for detecting various liver, kidney, and biliary tract disorders. Hence, precise quantification of this prognostic indicator holds significant value. Results : In this work, we have developed an ensemble‐based sensor (i.e., A3‐HSA), which successfully detects bilirubin in both buffer and urine samples. Two potential ensembles were initially tested: the A3‐HSA and A4‐HSA. While the former showed a turn‐off signal, the latter's signal remained largely unperturbed. Consequently, A3‐HSA conjugate was chosen for bilirubin sensing based on its distinct signal, and the limit of detection was found to be 2.77 µM. Molecular docking, MD simulations, and MMGBSA calculations were also performed to develop an understanding of the interactions between the ligands A3/A4/bilirubin and HSA. Calculations showed the strongest binding between bilirubin and HSA, thus facilitating the replacement of A3 from the IB site of HSA. The developed method was also validated using human urine samples and showed efficacy in measuring bilirubin levels at lower concentrations. Significance : The simplicity of the devised ensemble and its capability to work in urine allow for the easy detection of bilirubin, which is generally implicated in multiple ailments.
Key Findings
1
A3-HSA produced a distinct fluorescence turn-off response to bilirubin, whereas A4-HSA showed little signal change.
2
An HSA-mediated A3 ensemble sensor was developed for bilirubin detection in buffer and urine samples.
3
Docking, molecular dynamics, and MMGBSA calculations indicated that bilirubin binds HSA more strongly than A3 or A4, promoting A3 displacement from the IB site.
4
The A3-HSA sensor achieved a bilirubin detection limit of 2.77 µM.
5
The method was validated in human urine and measured bilirubin effectively at lower concentrations, offering a simple sensing approach.
Research Object
A3-HSA ensemble-based bilirubin sensor in buffer and urine samples
Research Subject
Photophysical sensing performance and HSA-mediated molecular recognition of bilirubin, including signal response, detection limit, and ligand displacement interactions