Microbial contamination in dental unit waterlines
Микробное загрязнение водопроводящих линий стоматологических установок
2003-06-01
SCID: 54.1/wsuaq2hn
Discuss with AI
biofilm formationcolony-forming units (CFU)dental unit waterlinesmicrobial contaminationpour-plate technique
Figures from the paper
Abstract (AI)
The quality of water in a dental unit is of considerable importance because patients and dental staff are regularly exposed to water and aerosol generated from the dental unit. The aim of this study was to evaluate the occurrence of microbial contamination in dental unit waterlines. Water samples were collected aseptically from the waterlines (reservoir, triple-syringe, high-speed) of 15 dental units. After serial dilution to 1:10(6) in APHA, the samples were seeded by the pour-plate technique and cultured in plate count agar (Difco) for 48 h at 32 degrees C. Analysis was based on the number of colony forming units (CFU). The Wilcoxon non-parametric test indicated that the levels of water contamination were highest in the triple-syringe (13 of 15) and in the high-speed (11 of 15); both levels were higher than those of the water reservoir. There was no significant statistical difference between the level of contamination in the triple-syringe and the high-speed as determined by the Mann-Whitney test [p(H0) = 40.98%; Z = - 0.2281]. Because biofilm forms on solid surfaces constantly bathed by liquid where microorganisms are present, these results indicate that the water in the dental unit may be contaminated by biofilm that forms in these tubules.
Key Findings
1
Both instrument waterlines had higher contamination than the water reservoirs, based on Wilcoxon non-parametric analysis.
2
Contamination levels were highest in triple-syringe waterlines, affecting 13 of 15 units, followed by high-speed waterlines, affecting 11 of 15.
3
Microbial contamination was evaluated in waterlines from the reservoirs, triple syringes, and high-speed instruments of 15 dental units using CFU counts.
4
The findings indicate that dental-unit water may be contaminated by biofilms forming inside tubules continuously exposed to water containing microorganisms.
5
Triple-syringe and high-speed waterlines showed no significant difference in contamination levels according to the Mann–Whitney test (p(H0) = 40.98%; Z = -0.2281).
Research Object
Dental unit waterlines, including the reservoir, triple-syringe, and high-speed lines
Research Subject
Occurrence and comparative levels of microbial contamination, including biofilm-associated contamination, in dental unit waterlines
Publication Details
Publication Date
2003-06-01
Journal
Publisher
ISSN