Nocturnal Glucose Control with an Artificial Pancreas at a Diabetes Camp

Ночной контроль уровня глюкозы с помощью искусственной поджелудочной железы в диабетическом лагере
Moshe Phillip, Tadej Battelino, Eran Atlas, Olga Kordonouri, Nataša Bratina, Shahar Miller, Torben Biester, Magdalena Avbelj Stefanija, Ido Muller, Revital Nimri, Thomas Danne
2013-02-27

artificial pancreasnocturnal glucose controlnocturnal hypoglycemiasensor-augmented insulin pumptype 1 diabetes
BACKGROUND: Recent studies have shown that an artificial-pancreas system can improve glucose control and reduce nocturnal hypoglycemia. However, it is not known whether such results can be replicated in settings outside the hospital. METHODS: In this multicenter, multinational, randomized, crossover trial, we assessed the short-term safety and efficacy of an artificial pancreas system for control of nocturnal glucose levels in patients (10 to 18 years of age) with type 1 diabetes at a diabetes camp. In two consecutive overnight sessions, we randomly assigned 56 patients to receive treatment with an artificial pancreas on the first night and a sensor-augmented insulin pump (control) on the second night or to the reverse order of therapies on the first and second nights. Thus, all the patients received each treatment in a randomly assigned order. The primary end points were the number of hypoglycemic events (defined as a sensor glucose value of <63 mg per deciliter [3.5 mmol per liter] for at least 10 consecutive minutes), the time spent with glucose levels below 60 mg per deciliter (3.3 mmol per liter), and the mean overnight glucose level for individual patients. RESULTS: On nights when the artificial pancreas was used, versus nights when the sensor-augmented insulin pump was used, there were significantly fewer episodes of nighttime glucose levels below 63 mg per deciliter (7 vs. 22) and significantly shorter periods when glucose levels were below 60 mg per deciliter (P=0.003 and P=0.02, respectively, after adjustment for multiplicity). Median values for the individual mean overnight glucose levels were 126.4 mg per deciliter (interquartile range, 115.7 to 139.1 [7.0 mmol per liter; interquartile range, 6.4 to 7.7]) with the artificial pancreas and 140.4 mg per deciliter (interquartile range, 105.7 to 167.4 [7.8 mmol per liter; interquartile range, 5.9 to 9.3]) with the sensor-augmented pump. No serious adverse events were reported. CONCLUSIONS: Patients at a diabetes camp who were treated with an artificial-pancreas system had less nocturnal hypoglycemia and tighter glucose control than when they were treated with a sensor-augmented insulin pump. (Funded by Sanofi and others; ClinicalTrials.gov number, NCT01238406.).
1
In a randomized crossover trial of 56 adolescents at a diabetes camp, an artificial pancreas was compared with a sensor-augmented insulin pump overnight.
2
Median individual mean overnight glucose was lower with the artificial pancreas than the control pump: 126.4 versus 140.4 mg/dL.
3
No serious adverse events were reported, supporting short-term safety in this nonhospital setting.
4
The artificial pancreas reduced nighttime glucose episodes below 63 mg/dL from 22 to 7 compared with the sensor-augmented pump.
5
The artificial pancreas significantly shortened the time spent below 60 mg/dL, with multiplicity-adjusted P=0.02.

Artificial-pancreas system used for nocturnal glucose management in 10- to 18-year-old patients with type 1 diabetes at a diabetes camp

Short-term safety and efficacy for reducing nocturnal hypoglycemia and controlling overnight glucose levels compared with a sensor-augmented insulin pump

Publication Details
Publication Date
2013-02-27
Journal
New England Journal of Medicine
Publisher
Massachusetts Medical Society
ISSN
0028-4793
Cited by
454
Access Type
Author Information
Authors
Moshe Phillip
Tadej Battelino
Eran Atlas
Olga Kordonouri
Nataša Bratina
Shahar Miller
Torben Biester
Magdalena Avbelj Stefanija
Ido Muller
Revital Nimri
Thomas Danne
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%