Effectiveness of Sensor-Augmented Insulin-Pump Therapy in Type 1 Diabetes
Эффективность терапии с помощью инсулиновой помпы с сенсором глюкозы при сахарном диабете 1-го типа
2010-06-29
SCID: 54.1/jkb3wxwp
Discuss with AI
glycated hemoglobin (HbA1c)multiple daily insulin injectionssensor-augmented pump therapysevere hypoglycemia ratetype 1 diabetes
Figures from the paper
Abstract (AI)
BACKGROUND: Recently developed technologies for the treatment of type 1 diabetes mellitus include a variety of pumps and pumps with glucose sensors. METHODS: In this 1-year, multicenter, randomized, controlled trial, we compared the efficacy of sensor-augmented pump therapy (pump therapy) with that of a regimen of multiple daily insulin injections (injection therapy) in 485 patients (329 adults and 156 children) with inadequately controlled type 1 diabetes. Patients received recombinant insulin analogues and were supervised by expert clinical teams. The primary end point was the change from the baseline glycated hemoglobin level. RESULTS: At 1 year, the baseline mean glycated hemoglobin level (8.3% in the two study groups) had decreased to 7.5% in the pump-therapy group, as compared with 8.1% in the injection-therapy group (P<0.001). The proportion of patients who reached the glycated hemoglobin target (<7%) was greater in the pump-therapy group than in the injection-therapy group. The rate of severe hypoglycemia in the pump-therapy group (13.31 cases per 100 person-years) did not differ significantly from that in the injection-therapy group (13.48 per 100 person-years, P=0.58). There was no significant weight gain in either group. CONCLUSIONS: In both adults and children with inadequately controlled type 1 diabetes, sensor-augmented pump therapy resulted in significant improvement in glycated hemoglobin levels, as compared with injection therapy. A significantly greater proportion of both adults and children in the pump-therapy group than in the injection-therapy group reached the target glycated hemoglobin level. (Funded by Medtronic and others; ClinicalTrials.gov number, NCT00417989.)
Key Findings
1
A significantly greater proportion of adults and children using pump therapy reached the target HbA1c level (<7%) compared with injection therapy.
2
In a 1-year randomized controlled trial of 485 patients, sensor-augmented insulin-pump therapy reduced mean HbA1c from 8.3% to 7.5%, versus 8.1% with multiple daily injections (P<0.001).
3
Rates of severe hypoglycemia did not differ significantly between pump therapy (13.31 cases per 100 person-years) and injection therapy (13.48 per 100 person-years; P=0.58).
4
There was no significant weight gain in either the pump-therapy group or the injection-therapy group over one year.
Research Object
Sensor-augmented insulin-pump therapy versus multiple daily insulin injections in patients with inadequately controlled type 1 diabetes
Research Subject
Effectiveness measured by change in glycated hemoglobin (HbA1c), proportion achieving target HbA1c (<7%), rates of severe hypoglycemia, and weight change over 1 year