Reducing risk of closed loop control of blood glucose in artificial pancreas using fractional calculus
Снижение риска замкнутого управления уровнем глюкозы в крови в искусственной поджелудочной железе с использованием дробного исчисления
2014-08-01
SCID: 54.1/9qcs6tut
Discuss with AI
artificial pancreasblood glucose controlfractional calculusmodel predictive controlnonlinear optimal control
Figures from the paper
Abstract (AI)
Healthcare costs in the US are among the highest in the world. Chronic diseases such as diabetes significantly contribute to these extensive costs. Despite technological advances to improve sensing and actuation devices, we still lack a coherent theory that facilitates the design and optimization of efficient and robust medical cyber-physical systems for managing chronic diseases. In this paper, we propose a mathematical model for capturing the complex dynamics of blood glucose time series (e.g., time dependent and fractal behavior) observed in real world measurements via fractional calculus concepts. Building upon our time dependent fractal model, we propose a novel model predictive controller for an artificial pancreas that regulates insulin injection. We verify the accuracy of our controller by comparing it to conventional non-fractal models using real world measurements and show how the nonlinear optimal controller based on fractal calculus concepts is superior to non-fractal controllers in terms of average risk index and prediction accuracy.
Key Findings
1
A fractional-calculus model captures time-dependent and fractal dynamics observed in real-world blood-glucose time series.
2
The fractional-calculus controller reduces average risk index compared with non-fractal controllers, indicating safer closed-loop glucose control.
3
The paper develops a model predictive controller for artificial-pancreas insulin regulation based on the fractional glucose model.
4
Using real-world measurements, the proposed nonlinear controller achieves better prediction accuracy than conventional non-fractal controllers.
Research Object
artificial pancreas closed-loop blood glucose control system
Research Subject
fractional-calculus-based modeling and model-predictive insulin regulation, focusing on blood-glucose dynamics, prediction accuracy, and risk reduction
Publication Details
Publication Date
2014-08-01
Journal
Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Publisher
Institute of Electrical and Electronics Engineers
ISSN
2375-7477
Cited by
18
Access Type
Author Information
Download PDF
Subscribe to digest
References available in scid.ai4
Model Predictive Control of Type 1 Diabetes: An in Silico Trial2007
Closed-Loop Artificial Pancreas Using Subcutaneous Glucose Sensing and Insulin Delivery and a Model Predictive Control Algorithm: The Virginia Experience2009
A Closed-Loop Artificial Pancreas Using Model Predictive Control and a Sliding Meal Size Estimator2009
Closed-Loop Artificial Pancreas Using Subcutaneous Glucose Sensing and Insulin Delivery and a Model Predictive Control Algorithm: Preliminary Studies in Padova and Montpellier2009