A Closed-Loop Artificial Pancreas Using Model Predictive Control and a Sliding Meal Size Estimator
Замкнутая система искусственной поджелудочной железы с использованием прогностического управления на основе модели и скользящего оценивателя размера приема пищи
2009-09-01
SCID: 54.1/cqd4xrw5
Discuss with AI
closed-loop artificial pancreashypoglycemia preventioninsulin-on-board constraintsmeal size estimationmodel predictive control
Figures from the paper
Abstract (AI)
The objective of this article is to present a comprehensive strategy for a closed-loop artificial pancreas. A meal detection and meal size estimation algorithm is developed for situations in which the subject forgets to provide a meal insulin bolus. A pharmacodynamic model of insulin action is used to provide insulin-on-board constraints to explicitly include the future effect of past and currently delivered insulin boluses. In addition, a supervisory pump shut-off feature is presented to avoid hypoglycemia. All of these components are used in conjunction with a feedback control algorithm using model predictive control (MPC). A model for MPC is developed based on a study of 20 subjects and is tested in a hypothetical clinical trial of 100 adolescent and 100 adult subjects using a Food and Drug Administration-approved diabetic subject simulator. In addition, a performance comparison of previously and newly proposed meal size estimation algorithms using 200 in silico subjects is presented. Using the new meal size estimation algorithm, the integrated artificial pancreas system yielded a daily mean glucose of 138 and 132 mg/dl for adolescents and adults, respectively, which is a substantial improvement over the MPC-only case, which yielded 159 and 145 mg/dl.
Key Findings
1
A comprehensive closed-loop artificial pancreas integrates meal detection, meal-size estimation, insulin-on-board constraints, supervisory pump shut-off, and MPC feedback control.
2
A pharmacodynamic insulin-action model explicitly constrains insulin-on-board, accounting for future effects of previously and currently delivered insulin.
3
In simulator trials, the integrated system achieved daily mean glucose levels of 138 mg/dl in adolescents and 132 mg/dl in adults.
4
The integrated system substantially outperformed MPC-only control, which produced mean glucose levels of 159 mg/dl in adolescents and 145 mg/dl in adults.
5
The meal detection and size estimation algorithm addresses missed meal boluses, enabling the system to compensate when subjects fail to announce meals.
Research Object
closed-loop artificial pancreas system for insulin-dependent diabetes management
Research Subject
glucose-control performance and hypoglycemia prevention with meal detection and size estimation, insulin-on-board constraints, and MPC-based insulin delivery
Publication Details
Publication Date
2009-09-01
Journal
Journal of Diabetes Science and Technology
Publisher
SAGE Publishing
ISSN
1932-2968
Cited by
165
Open access PDF
Access Type
Author Information
Download PDF
Subscribe to digest
References available in scid.ai5
Quantitative estimation of insulin sensitivity.1979
In Silico Preclinical Trials: A Proof of Concept in Closed-Loop Control of Type 1 Diabetes2009
Model Predictive Control of Type 1 Diabetes: An in Silico Trial2007
Safety Constraints in an Artificial Pancreatic β Cell: An Implementation of Model Predictive Control with Insulin on Board2009
In Silico Preclinical Trials: Methodology and Engineering Guide to Closed-Loop Control in Type 1 Diabetes Mellitus2009
Cited by6
Closed-loop insulin delivery for treatment of type 1 diabetes2011
The Artificial Pancreas and Meal Control: An Overview of Postprandial Glucose Regulation in Type 1 Diabetes2018
Model-Based Detection and Classification of Insulin Pump Faults and Missed Meal Announcements in Artificial Pancreas Systems for Type 1 Diabetes Therapy2020
Reducing risk of closed loop control of blood glucose in artificial pancreas using fractional calculus2014
A model-based approach to synthesizing insulin infusion pump usage parameters for diabetic patients2012
A Wearable Closed-Loop Insulin Delivery System Based on Low-Power SoCs2019