Resonance sonomanometry for noninvasive, continuous monitoring of blood pressure

Резонансная сономанометрия для неинвазивного непрерывного мониторинга артериального давления
Yaser S. Abu‐Mostafa, Matthew Fu, William P. Dempsey, Raymond Jimenez, Dominic Yurk, Steven J. Dell, Austin C. Rutledge, Aditya Rajagopal, Alaina Brinley Rajagopal
2024-06-28

arterial dimension imagingnoninvasive continuous blood pressureresonance sonomanometryultrasound arterial wall resonances
Cardiovascular disease is the leading cause of death worldwide. Existing methods for continuous, noninvasive blood pressure (BP) monitoring suffer from poor accuracy, uncomfortable form factors, or a need for frequent calibration, limiting their adoption. We introduce a new framework for continuous BP measurement that is noninvasive and calibration-free called resonance sonomanometry. The method uses ultrasound imaging to measure both the arterial dimensions and artery wall resonances that are induced by acoustic stimulation, which offers a direct measure of BP by a fully determined physical model. The approach and model are validated in vitro using arterial mock-ups and then in multiple arteries in human subjects. This approach offers the promise of robust continuous BP measurements, providing significant benefits for early diagnosis and treatment of cardiovascular disease.
1
Approach validated in vitro using arterial mock-ups and in human subjects across multiple arteries.
2
Artery wall resonances provide a direct measure of blood pressure via a fully determined physical model.
3
Introduces resonance sonomanometry: a noninvasive, continuous, and calibration-free framework for blood pressure measurement.
4
Method combines ultrasound imaging of arterial dimensions with measurement of artery wall resonances induced by acoustic stimulation.
5
Promised benefit: robust continuous BP measurements that could improve early diagnosis and treatment of cardiovascular disease.

Human arteries (and arterial mock-ups) measured by ultrasound in resonance sonomanometry for noninvasive continuous blood pressure monitoring

Measurement of arterial dimensions and artery wall resonances induced by acoustic stimulation to provide a direct, calibration-free estimate of blood pressure via a fully determined physical model

Publication Details
Publication Date
2024-06-28
Journal
Publisher
ISSN
Cited by
15
Access Type
Author Information
Authors
Yaser S. Abu‐Mostafa
Matthew Fu
William P. Dempsey
Raymond Jimenez
Dominic Yurk
Steven J. Dell
Austin C. Rutledge
Aditya Rajagopal
Alaina Brinley Rajagopal
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%