Voltammetric Sensing of Bilirubin Based on Nafion/Electrochemically Reduced Graphene Oxide Composite Modified Glassy Carbon Electrode
Вольтамперометрическое определение билирубина с использованием стеклоуглеродного электрода, модифицированного композитом на основе Нафиона и электрохимически восстановленного оксида графена
2015-02-23
SCID: 54.1/d7hj3bhu
Discuss with AI
BilirubinGlassy carbon electrodeHuman serum analysisNafion/electrochemically reduced graphene oxideSquare-wave voltammetry
Figures from the paper
Abstract (AI)
A sensitive square wave voltammetric sensor for the quantification of bilirubin (BR) at the Nafion-electrochemically reduced graphene oxide (NF/ER-GO) hybrid composite modified conventional glassy carbon electrode was offered. The voltammetric oxidation conducts of bilirubin at the NF/ER-GO modified GC electrode were examined in acetone containing H2SO4 as the supporting electrolyte. The electrochemical conduct of BR on NF/ER-GO hybrid modified GC electrodes was explored by electrochemical impedance spectroscopy, cyclic voltammetry and square-wave voltammetry (SWV). The oxidation peak current response of BR achieved at the NF/ER-GO modified GC electrode was 3 times higher than that of the unmodified GC electrode. At specified conditions, the linear calibration curve for BR showed two good linear segments: the first linear segment increased from 2.0 μM to 20 μM and second linear segment increased up to 70 μM. The limit of detection was calculated as 0.84 μM using SWV. The results showed that the NF/ER-GO modified electrode can be applied to determine BR without interference from ascorbic acid and uric acid, while ensuring good sensitivity, selectivity, and reproducibility. Finally, the proposed sensor was successfully employed to detect bilirubin in human serum samples. Keywords: Bilirubin, voltammetry, graphene, biomolecule, serum analysis.
Key Findings
1
A Nafion/electrochemically reduced graphene oxide composite modified glassy carbon electrode was developed for square-wave voltammetric bilirubin quantification.
2
Bilirubin showed two linear calibration ranges, from 2.0–20 μM and from 20–70 μM, under optimized conditions.
3
The modified electrode produced a bilirubin oxidation peak current three times higher than that of an unmodified glassy carbon electrode.
4
The sensor achieved a square-wave voltammetric detection limit of 0.84 μM.
5
The sensor determined bilirubin in human serum with good sensitivity, selectivity, and reproducibility, without interference from ascorbic acid or uric acid.
Research Object
Bilirubin in human serum measured at a Nafion/electrochemically reduced graphene oxide composite-modified glassy carbon electrode
Research Subject
Voltammetric oxidation and quantitative sensing performance of bilirubin, including sensitivity, selectivity, linear response, detection limit, and reproducibility