A polishable amperometric biosensor for bilirubin
Полируемый амперометрический биосенсор для определения билирубина
1990-11-01
SCID: 54.1/c2yr74uj
Discuss with AI
amperometric biosensorbilirubin detectionbilirubin oxidasegraphite-epoxy electrodehorseradish peroxidase
Figures from the paper
Abstract (AI)
Abstract An amperometric enzyme electrode for bilirubin based on the incorporation of bilirubin oxidase and horseradish peroxidase within a graphite‐epoxy matrix is reported. The resulting bienzyme electrodes are renewable (by polishing) and offer low‐potential detection of bilirubin. The limit of detection is 4 × 10−6 M and the response is linear up to 1 × 10−4 M. The influence of various experimental variables is described. The surface‐to‐surface reproducibility is 14% (in terms of relative standard deviation). The fabrication scheme is simple and the response times are short. The prospects of using amperometric bilirubin electrodes for diagnostic work are discussed.
Key Findings
1
A renewable amperometric bilirubin biosensor incorporates bilirubin oxidase and horseradish peroxidase into a graphite–epoxy matrix.
2
Surface-to-surface reproducibility is 14% relative standard deviation, with simple fabrication and short response times.
3
The bienzyme electrode enables low-potential bilirubin detection with a limit of detection of 4 × 10−6 M.
4
The electrode can be renewed by polishing, supporting potential use in diagnostic bilirubin measurements.
5
The sensor response is linear for bilirubin concentrations up to 1 × 10−4 M.
Research Object
polishable bienzyme amperometric bilirubin electrode based on a graphite–epoxy matrix incorporating bilirubin oxidase and horseradish peroxidase
Research Subject
low-potential bilirubin detection performance, including detection limit, linear range, response time, reproducibility, and the influence of experimental variables
Publication Details
Publication Date
1990-11-01
Journal
Publisher
ISSN
Cited by
75
Access Type
Author Information
Download PDF
Subscribe to digest
Cited by6
Electrochemical biosensor with graphene oxide nanoparticles and polypyrrole interface for the detection of bilirubin2016
Determination of Bilirubin by Fiberoptic Biosensor1996
Nafion/Multi-wall Carbon Nanotubes Composite Modified Glassy Carbon Electrode for Sensitive Determination of Bilirubin2015
Bilirubin Biosensing by Using of Catalase and ZnS Nanoparticles as Modifier2016
Voltammetric Sensing of Bilirubin Based on Nafion/Electrochemically Reduced Graphene Oxide Composite Modified Glassy Carbon Electrode2015
On the Relationship between the Characteristics of Bilirubin Oxidases and O2 Cathodes Based on Their “Wiring”2002