A Wearable Closed-Loop Insulin Delivery System Based on Low-Power SoCs

Носимая система замкнутого управления доставкой инсулина на основе энергоэффективных однокристальных систем
Jesús Berián, Ignácio Bravo, Alfredo Gardel, José Luis Lázaro Galilea, Sergio Hernández
2019-05-31

UVa/Padova Simulatorclosed-loop insulin deliverycontinuous glucose monitoringlow-power SoCswearable embedded system
The number of patients living with diabetes has increased significantly in recent years due to several factors. Many of these patients are choosing to use insulin pumps for their treatment, artificial systems that administer their insulin and consist of a glucometer and an automatic insulin supply working in an open loop. Currently, only a few closed-loop insulin delivery devices are commercially available. The most widespread systems among patients are what have been called the “Do-It-Yourself Hybrid Closed-Loop systems.” These systems require the use of platforms with high computing power. In this paper, we will present a novel wearable system for insulin delivery that reduces the energy and computing consumption of the platform without affecting the computation requirements. Patients’ information is obtained from a commercial continuous glucose sensor and a commercial insulin pump operating in a conventional manner. An ad-hoc embedded system will connect with the pump and the sensor to collect the glucose data and process it. That connection is accomplished through a radiofrequency channel that provides a suitable system for the patient. Thus, this system does not require to be connected to any other processor, which increases the overall stability. Using parameters configured by the patient, the control system will make automatic adjustments in the basal insulin infusion thereby bringing the patient’s glycaemia to the target set by a doctor’s prescription. The results obtained will be satisfactory as long as the configured parameters faithfully match the specific characteristics of the patient. Results from the simulation of 30 virtual patients (10 adolescents, 10 adults, and 10 children), using a python implementation of the FDA-approved (Food and Drug Administration) UVa (University of Virginia)/Padova Simulator and a python implementation of the proposed algorithm, are presented.
1
A novel wearable closed-loop insulin delivery system uses low-power SoCs to reduce energy and computing consumption without changing the algorithm’s computational requirements.
2
Simulation with 30 virtual patients—10 adolescents, 10 adults, and 10 children—used the FDA-approved UVA/Padova Simulator and a Python implementation of the proposed algorithm.
3
The controller automatically adjusts basal insulin infusion using patient-configured parameters to drive glucose toward a physician-prescribed target.
4
The embedded system wirelessly connects a commercial continuous glucose monitor and insulin pump, eliminating dependence on an external high-performance processor and improving system stability.
5
The system achieved satisfactory simulated results when its configured parameters accurately matched each patient’s individual characteristics.

a wearable closed-loop insulin delivery system based on low-power SoCs, integrating a continuous glucose sensor and an insulin pump

low-power autonomous closed-loop control of basal insulin infusion to regulate patients’ glycaemia toward a prescribed target

Publication Details
Publication Date
2019-05-31
Journal
Electronics
Publisher
Multidisciplinary Digital Publishing Institute
ISSN
2079-9292
Cited by
12
Access Type
Author Information
Authors
Jesús Berián
Ignácio Bravo
Alfredo Gardel
José Luis Lázaro Galilea
Sergio Hernández
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%