Metal Artifact Reduction With MAVRIC SL at 3-T MRI in Patients With Hip Arthroplasty
Уменьшение металлических артефактов с помощью MAVRIC SL при МРТ на аппарате 3 Тл у пациентов с эндопротезированием тазобедренного сустава
2014-12-24
SCID: 54.1/mytd6w49
Discuss with AI
3-T MRIMAVRIC SLfast spin-echohip arthroplastymetal artifact reduction
Figures from the paper
Abstract (AI)
OBJECTIVE: The objective of our study was to compare the multiacquisition variable-resonance image combination selective (MAVRIC SL) sequence with the 2D fast spin-echo (FSE) sequence for metal artifact reduction on 3-T MRI in patients with hip arthroplasty (HA). MATERIALS AND METHODS: Matched 2D FSE and MAVRIC SL images of 21 hips (19 patients with HA) were included in the study group. Paired image sets, composed of 13 coronal and 12 axial slices (total, 25 image sets), of the 21 hips were evaluated. For quantitative analysis, the artifact area was measured at the level of the hip and femur. For qualitative analysis, two musculoskeletal radiologists independently compared paired 2D FSE and MAVRIC SL sets in terms of artifacts, depiction of anatomic detail, level of diagnostic confidence, and detection of abnormal findings. RESULTS: The measured artifact area was significantly smaller (p < 0.05) on MAVRIC SL than 2D FSE at both the level of hip (59.9% reduction with MAVRIC SL) and femur (31.3% reduction with MAVRIC SL). The artifact score was also significantly decreased (p < 0.0001) with MAVRIC SL compared with 2D FSE for both reviewers. The hip joint capsule and the muscle and tendon attachment sites of the obturator externus and iliopsoas muscles were better depicted with MAVRIC SL than 2D FSE (p < 0.0125). Abnormal findings were significantly better shown on MAVRIC SL imaging compared with 2D FSE imaging (p < 0.0001). CONCLUSION: The MAVRIC SL sequence can significantly reduce metal artifact on 3-T MRI compared with the 2D FSE sequence and can increase diagnostic confidence of 3-T MRI in patients with total HA.
Key Findings
1
Abnormal findings were significantly better detected with MAVRIC SL, increasing diagnostic confidence in 3-T MRI after total hip arthroplasty.
2
Artifact area decreased by 59.9% at the hip and 31.3% at the femur with MAVRIC SL versus 2D FSE.
3
Both radiologists assigned significantly lower artifact scores to MAVRIC SL images than to 2D FSE images.
4
MAVRIC SL provided better visualization of the hip capsule and obturator externus and iliopsoas muscle-tendon attachment sites.
5
MAVRIC SL significantly reduced metal artifact area compared with 2D FSE in 3-T MRI of hip arthroplasty patients.
Research Object
Metal artifacts on 3-T MRI images of patients with total hip arthroplasty
Research Subject
Metal artifact reduction and improvement of anatomic depiction, diagnostic confidence, and abnormal-finding detection using MAVRIC SL compared with 2D FSE
Publication Details
Publication Date
2014-12-24
Journal
Publisher
ISSN
Access Type
Author Information
Download PDF
Subscribe to digest