Nafion/Multi-wall Carbon Nanotubes Composite Modified Glassy Carbon Electrode for Sensitive Determination of Bilirubin
Композитно-модифицированный стеклоуглеродный электрод на основе Нафиона/многостенных углеродных нанотрубок для чувствительного определения билирубина
2015-06-03
SCID: 54.1/qp6s544d
Discuss with AI
Bilirubin detectionElectrochemical sensorGlassy carbon electrodeNafion/MWCNT compositeSquare-wave voltammetry
Figures from the paper
Abstract (AI)
A composite picture of multi-wall carbon nanotubes (MWCNTs) with Nafion (NF) cation exchange membrane is prepared by cast deposition on glassy carbon electrodes. A sensitive electrochemical sensor for the detection of bilirubin (BR) at the NF/MWCNTs hybrid modified glassy carbon electrode was offered. The electrochemical behaviors of bilirubin at the NF/MWCNTs hybrid electrode were studied in acetone containing sulfuric acid as the supporting electrolyte. Compared with bare glassy carbon electrode (GCE), the proposed integrated sensor, improved analytical performance characteristics in the electrochemical oxidation of bilirubin. The electrochemistry of bilirubin on NF/MWCNTs hybrid electrodes were studied by electrochemical impedance spectroscopy, cyclic voltammetry (CV) and square-wave voltammetry (SWV). The oxidation peak current of bilirubin obtained at the NF/MWCNTs modified GC electrode was 10 times higher than that of the unmodified glassy carbon electrode. To optimize the experimental conditions, a linear calibration plot for bilirubin was obtained between 0.8 µM and 10 µM with the detection limit of 0.014 µM. Finally, the proposed sensor was successfully employed to determine bilirubin in human urine samples.
Key Findings
1
A Nafion/multi-wall carbon nanotube composite was deposited on glassy carbon to create a bilirubin-sensitive electrochemical sensor.
2
Bilirubin showed a linear electrochemical response from 0.8 to 10 µM, with a detection limit of 0.014 µM.
3
Electrochemical impedance spectroscopy, cyclic voltammetry, and square-wave voltammetry were used to characterize bilirubin behavior at the hybrid electrode.
4
The hybrid-modified electrode enhanced bilirubin oxidation, producing a peak current 10 times higher than that of bare glassy carbon.
5
The sensor was successfully applied to bilirubin determination in human urine samples.
Research Object
Nafion/multi-wall carbon nanotube composite-modified glassy carbon electrode for bilirubin detection
Research Subject
Electrochemical oxidation behavior and sensitive analytical determination of bilirubin, including the sensor’s response and detection performance
Publication Details
Publication Date
2015-06-03
Journal
Publisher
ISSN
Cited by
9