Closed-Loop Insulin Delivery During Pregnancy Complicated by Type 1 Diabetes
Замкнутая система доставки инсулина во время беременности, осложнённой сахарным диабетом 1-го типа
2011-01-08
SCID: 54.1/v7arqpgg
Discuss with AI
closed-loop insulin deliverycontinuous glucose monitoringmodel predictive controlpregnancytype 1 diabetes
Figures from the paper
Abstract (AI)
OBJECTIVE: This study evaluated closed-loop insulin delivery with a model predictive control (MPC) algorithm during early (12-16 weeks) and late gestation (28-32 weeks) in pregnant women with type 1 diabetes. RESEARCH DESIGN AND METHODS: Ten women with type 1 diabetes (age 31 years, diabetes duration 19 years, BMI 24.1 kg/m(2), booking A1C 6.9%) were studied over 24 h during early (14.8 weeks) and late pregnancy (28.0 weeks). A nurse adjusted the basal insulin infusion rate from continuous glucose measurements (CGM), fed into the MPC algorithm every 15 min. Mean glucose and time spent in target (63-140 mg/dL), hyperglycemic (>140 to ≥ 180 mg/dL), and hypoglycemic (<63 to ≤ 50 mg/dL) were calculated using plasma and sensor glucose measurements. Linear mixed-effects models were used to compare glucose control during early and late gestation. RESULTS: During closed-loop insulin delivery, median (interquartile range) plasma glucose levels were 117 (100.8-154.8) mg/dL in early and 126 (109.8-140.4) mg/dL in late gestation (P = 0.72). The overnight mean (interquartile range) plasma glucose time in target was 84% (50-100%) in early and 100% (94-100%) in late pregnancy (P = 0.09). Overnight mean (interquartile range) time spent hyperglycemic (>140 mg/dL) was 7% (0-40%) in early and 0% (0-6%) in late pregnancy (P = 0.25) and hypoglycemic (<63 mg/dL) was 0% (0-3%) and 0% (0-0%), respectively (P = 0.18). Postprandial glucose control, glucose variability, insulin infusion rates, and CGM sensor accuracy were no different in early or late pregnancy. CONCLUSIONS: MPC algorithm performance was maintained throughout pregnancy, suggesting that overnight closed-loop insulin delivery could be used safely during pregnancy. More work is needed to achieve optimal postprandial glucose control.
Key Findings
1
A model predictive control closed-loop system maintained comparable glucose control during early and late pregnancy in women with type 1 diabetes.
2
Median plasma glucose was 117 mg/dL in early gestation and 126 mg/dL in late gestation, with no significant difference (P = 0.72).
3
Overnight time in target was 84% during early pregnancy and 100% during late pregnancy, while hypoglycemia remained minimal in both periods.
4
Postprandial control, glucose variability, insulin infusion rates, and CGM sensor accuracy did not differ between early and late gestation.
5
The findings support safe overnight closed-loop insulin delivery throughout pregnancy, although postprandial glucose control requires further improvement.
Research Object
closed-loop insulin delivery using a model predictive control algorithm in pregnant women with type 1 diabetes during early and late gestation
Research Subject
maintenance of glucose control, safety, and MPC performance across gestation, including overnight and postprandial glycemic outcomes, glucose variability, insulin infusion requirements, and sensor accuracy