Validation and Subgroup Analysis of the Accuracy of the Photoplethysmography-based Microlife Cuffless Upper-arm Wearable Blood Pressure Monitor

Валидация и анализ подгрупп точности безманжетного носимого тонометра для верхней конечности Microlife на основе фотоплетизмографии
Bo‐Wei Chen, Hanlin Wang, Ching-Fu Wang, Pei-Hsin Kuo, Shih-Zhang Li, Ssu‐Ju Li, You‐Yin Chen, Sheng-Huang Lin
2025-12-11

Microlife cuffless upper-arm wearableaccuracy across varying skin tonesaccuracy in elderly individualsphotoplethysmography-based cuffless blood pressure monitoringvalidation and subgroup analysis
BACKGROUND: This study (1) validates the accuracy of the photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure (BP) monitor according to the AAMI/ESH/ISO 81060-2:2018/Amd 1:2020 standard; (2) investigates the device's performance across age groups and skin tone; and (3) evaluates the advantages of cuffless BP monitoring in terms of estimation variability, required time, and stability. METHODS: A total of 120 participants aged ≥20 years were recruited from the general population, ensuring the representation of diverse age and skin tone groups. The device was calibrated against a clinically validated mercury sphygmomanometer. Sequential same-arm BP estimation were performed. Accuracy was analyzed according to the ISO criteria, and further subgroup analysis was conducted to compare results between participants aged ≥65 and <65 years, as well as between participants with lighter and darker skin tones (classified according to the Fitzpatrick system). Additionally, estimation stability and variability were evaluated using 3 averaged BP estimations. RESULTS: The results demonstrated that the Microlife device met the AAMI/ESH/ISO accuracy criteria. Subgroup analysis revealed consistent accuracy across age and skin tone groups, with slight differences warranting further exploration. Notably, the cuffless design enabled faster estimations with reduced variation between the 3 averaged readings, showcasing its potential for home BP monitoring and frequent self-assessments. CONCLUSIONS: These findings support the clinical potential of photoplethysmography-based cuffless BP monitoring in diverse adult populations, particularly for facilitating rapid and stable BP estimations in elderly individuals and users with varying skin tones. Further large-scale studies are warranted to corroborate and build upon these observations.
1
Authors recommend further large-scale studies to corroborate and extend these observations.
2
Device facilitates rapid and stable blood pressure estimations in elderly individuals.
3
Device produces reliable blood pressure measurements for users with varying skin tones.
4
Photoplethysmography-based cuffless Microlife upper-arm wearable provides clinically promising blood pressure monitoring across diverse adult populations.

Photoplethysmography-based Microlife cuffless upper-arm wearable blood pressure monitor

Accuracy validation and subgroup analysis of blood pressure estimation performance across diverse adult populations (including elderly individuals and varying skin tones)

Publication Details
Publication Date
2025-12-11
Journal
Publisher
ISSN
Cited by
0
Access Type
Author Information
Authors
Bo‐Wei Chen
Hanlin Wang
Ching-Fu Wang
Pei-Hsin Kuo
Shih-Zhang Li
Ssu‐Ju Li
You‐Yin Chen
Sheng-Huang Lin
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%