Evaluation of a novel cuffless photoplethysmography-based wristband for measuring blood pressure according to the regulatory standards

Оценка нового бескабельного (безмногослойного манжеты) наручного датчика на основе фотоплетизмографии для измерения артериального давления в соответствии с нормативными стандартами
Stefan H. J. Monnink, Eelko Ronner, Mariska van Vliet, Mathijs J. Kuiper, Jan Constandse, Dieke Hoftijzer
2024-02-08

AAMI/ESH/ISO Universal StandardPPG-guided blood pressure algorithmarterial sub-clavian referencecuffless photoplethysmography wristbandmachine learning BP algorithm
Aims: Elevated blood pressure (BP) is a key risk factor in cardiovascular diseases. However, obtaining reliable and reproducible BP remains a challenge. This study, therefore, aimed to evaluate a novel cuffless wristband, based on photoplethysmography (PPG), for continuous BP monitoring. Methods and results: Predictions by a PPG-guided algorithm were compared to arterial BP measurements (in the sub-clavian artery), obtained during cardiac catheterization. Eligible patients were included and screened based on AAMI/European Society of Hypertension (ESH)/ISO Universal Standard requirements. The machine learning-based BP algorithm required three cuff-based initialization measurements in combination with ∼100 features (signal-derived and patient demographic-based). Ninety-seven patients and 420 samples were included. Mean age, weight, and height were 67.1 years (SD 11.1), 83.4 kg (SD 16.1), and 174 cm (SD 10), respectively. Systolic BP was ≤100 mmHg in 48 samples (11%) and ≥160 mmHg in 106 samples (25%). Diastolic BP was ≤70 mmHg in 222 samples (53%) and ≥85 mmHg in 99 samples (24%). The algorithm showed mean errors of ±3.7 mmHg (SD 4.4 mmHg) and ±2.5 mmHg (SD 3.7 mmHg) for systolic and diastolic BP, respectively. Similar results were observed across all genders and skin colours (Fitzpatrick I-VI). Conclusion: This study provides initial evidence for the accuracy of a PPG-based BP algorithm in combination with a cuffless wristband across a range of BP distributions. This research complies with the AAMI/ESH/ISO Universal Standard, however, further research is required to evaluate the algorithms performance in light of the remaining European Society of Hypertension recommendations. Clinical trial registration: www.clinicaltrials.gov, NCT05566886.
1
A novel cuffless PPG-based wristband with a machine-learning algorithm was evaluated against invasive sub-clavian arterial BP measurements during cardiac catheterization.
2
In 97 patients (420 samples) the algorithm achieved mean errors of +/−3.7 mmHg (SD 4.4) for systolic BP and +/−2.5 mmHg (SD 3.7) for diastolic BP.
3
Performance was consistent across genders and Fitzpatrick skin types I–VI.
4
The algorithm required three cuff-based initialization measurements plus ~100 signal-derived and demographic features for BP prediction.
5
The study meets AAMI/ESH/ISO Universal Standard requirements, but further research is needed to address remaining European Society of Hypertension recommendations.

Novel cuffless photoplethysmography-based wristband for continuous blood pressure monitoring

Accuracy and agreement of the PPG-based blood pressure algorithm/wristband compared to invasive arterial (sub-clavian) blood pressure measurements, evaluated against AAMI/ESH/ISO Universal Standard criteria across BP ranges and demographics

Publication Details
Publication Date
2024-02-08
Journal
Publisher
ISSN
Cited by
24
Access Type
Author Information
Authors
Stefan H. J. Monnink
Eelko Ronner
Mariska van Vliet
Mathijs J. Kuiper
Jan Constandse
Dieke Hoftijzer
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat
Make a presentation
100%