Evaluation of a novel cuffless photoplethysmography-based wristband for measuring blood pressure according to the regulatory standards
Оценка нового бескабельного (безмногослойного манжеты) наручного датчика на основе фотоплетизмографии для измерения артериального давления в соответствии с нормативными стандартами
2024-02-08
SCID: 54.1/mkd6rux9
Discuss with AI
AAMI/ESH/ISO Universal StandardPPG-guided blood pressure algorithmarterial sub-clavian referencecuffless photoplethysmography wristbandmachine learning BP algorithm
Figures from the paper
Abstract (AI)
Aims: Elevated blood pressure (BP) is a key risk factor in cardiovascular diseases. However, obtaining reliable and reproducible BP remains a challenge. This study, therefore, aimed to evaluate a novel cuffless wristband, based on photoplethysmography (PPG), for continuous BP monitoring. Methods and results: Predictions by a PPG-guided algorithm were compared to arterial BP measurements (in the sub-clavian artery), obtained during cardiac catheterization. Eligible patients were included and screened based on AAMI/European Society of Hypertension (ESH)/ISO Universal Standard requirements. The machine learning-based BP algorithm required three cuff-based initialization measurements in combination with ∼100 features (signal-derived and patient demographic-based). Ninety-seven patients and 420 samples were included. Mean age, weight, and height were 67.1 years (SD 11.1), 83.4 kg (SD 16.1), and 174 cm (SD 10), respectively. Systolic BP was ≤100 mmHg in 48 samples (11%) and ≥160 mmHg in 106 samples (25%). Diastolic BP was ≤70 mmHg in 222 samples (53%) and ≥85 mmHg in 99 samples (24%). The algorithm showed mean errors of ±3.7 mmHg (SD 4.4 mmHg) and ±2.5 mmHg (SD 3.7 mmHg) for systolic and diastolic BP, respectively. Similar results were observed across all genders and skin colours (Fitzpatrick I-VI). Conclusion: This study provides initial evidence for the accuracy of a PPG-based BP algorithm in combination with a cuffless wristband across a range of BP distributions. This research complies with the AAMI/ESH/ISO Universal Standard, however, further research is required to evaluate the algorithms performance in light of the remaining European Society of Hypertension recommendations. Clinical trial registration: www.clinicaltrials.gov, NCT05566886.
Key Findings
1
A novel cuffless PPG-based wristband with a machine-learning algorithm was evaluated against invasive sub-clavian arterial BP measurements during cardiac catheterization.
2
In 97 patients (420 samples) the algorithm achieved mean errors of +/−3.7 mmHg (SD 4.4) for systolic BP and +/−2.5 mmHg (SD 3.7) for diastolic BP.
3
Performance was consistent across genders and Fitzpatrick skin types I–VI.
4
The algorithm required three cuff-based initialization measurements plus ~100 signal-derived and demographic features for BP prediction.
5
The study meets AAMI/ESH/ISO Universal Standard requirements, but further research is needed to address remaining European Society of Hypertension recommendations.
Research Object
Novel cuffless photoplethysmography-based wristband for continuous blood pressure monitoring
Research Subject
Accuracy and agreement of the PPG-based blood pressure algorithm/wristband compared to invasive arterial (sub-clavian) blood pressure measurements, evaluated against AAMI/ESH/ISO Universal Standard criteria across BP ranges and demographics
Publication Details
Publication Date
2024-02-08
Journal
Publisher
ISSN
Cited by
24
Open access PDF
Access Type
Author Information
Download PDF
Subscribe to digest