This article sets forth guidelines for in silico (simulation-based) proof-of-concept testing of artificial pancreas control algorithms. The goal was to design a test procedure that can facilitate regulatory approval [e.g., Food and Drug Administration Investigational Device Exemption] for General Clinical Research Center experiments without preliminary testing on animals. The methodology is designed around a software package, based on a recent meal simulation model of the glucose-insulin system. Putting a premium on generality, this document starts by specifying a generic, rather abstract, meta-algorithm for control. The meta-algorithm has two main components: (1) patient assessment and tuning of control parameters, i.e., algorithmic processes for collection and processing patient data prior to closed-loop operation, and (2) controller warm-up and run-time operation, i.e., algorithmic processes for initializing controller states and managing blood glucose. The simulation-based testing methodology is designed to reveal the conceptual/mathematical operation of both main components, as applied to a large population of in silico patients with type 1 diabetes mellitus.
1
A software package implementing a recent meal simulation model of the glucose–insulin system underpins the testing procedure.
2
Testing across a large simulated population is designed to expose the conceptual and mathematical operation of both control components.
3
The article provides guidelines for simulation-based proof-of-concept testing of artificial pancreas control algorithms.
4
The methodology aims to support regulatory approval for clinical research experiments without preliminary animal testing.
5
The proposed meta-algorithm separates patient-specific assessment and parameter tuning from controller warm-up and runtime blood-glucose management.

Artificial pancreas closed-loop control for blood glucose management in in silico patients with type 1 diabetes mellitus

Simulation-based evaluation of patient assessment and controller tuning, initialization, and run-time operation across a large virtual patient population to characterize the conceptual and mathematical behavior of control algorithms

Publication Details
Publication Date
2009-03-01
Journal
Journal of Diabetes Science and Technology
Publisher
SAGE Publishing
ISSN
1932-2968
Cited by
92
Access Type
Author Information
Authors
Stephen D. Patek
B. Wayne Bequette
Marc D. Breton
Bruce A. Buckingham
Eyal Dassau
Francis J. Doyle
John W. Lum
Lalo Magni
Howard Zisser
Explore further
Open the scid.ai AI chat with a ready-made request: it will find papers on a similar topic and help build a literature review.
Find similar papers in the chat →
Make a presentation
100%