A model-based approach to synthesizing insulin infusion pump usage parameters for diabetic patients
Модельный подход к синтезу параметров использования инфузионной инсулиновой помпы для пациентов с диабетом
2012-10-01
SCID: 54.1/zq4jg36r
Discuss with AI
global optimizationinsulin infusion pumpsinsulin-glucose regulatory systeminsulin-to-carbohydrate ratiosmeal bolus calculation
Figures from the paper
Abstract (AI)
We present a model-based approach to synthesizing insulin infusion pump usage parameters against varying meal scenarios and physiological conditions. Insulin infusion pumps are commonly used by type-1 diabetic patients to control their blood glucose levels. The amounts of insulin to be infused are calculated based on parameters such as insulin-to-carbohydrate ratios and correction factors that need to be calibrated carefully for each patient. Frequent and careful calibration of these parameters is essential for avoiding complications such as hypoglycemia and hyperglycemia. In this paper, we propose to synthesize optimal parameters for meal bolus calculation starting from models of the patient's insulin-glucose regulatory system and the infusion pump. Various off-the-shelf global optimization techniques are used to search for parameter values that minimize a penalty function defined over the predicted glucose sensor readings. The penalty function “rewards” glucose levels that lie within the prescribed ranges and “penalizes” the occurrence of hypoglycemia and hyperglycemia. We evaluate our approach using a model of the insulin-glucose regulatory system proposed by Dalla Man et al. Using this model, we compare various strategies for optimizing pump usage parameters for a virtual population of in-silico patients.
Key Findings
1
A model-based method synthesizes patient-specific insulin pump parameters for meal bolus calculation under varying meals and physiological conditions.
2
Evaluation with the Dalla Man insulin-glucose model compares parameter-optimization strategies across a virtual population of in-silico patients.
3
The approach optimizes insulin-to-carbohydrate ratios and correction factors using global optimization over predicted glucose sensor readings.
4
The method aims to reduce risks associated with frequent parameter miscalibration, including hypoglycemia and hyperglycemia.
5
The penalty function rewards glucose values within prescribed ranges while penalizing hypoglycemia and hyperglycemia.
Research Object
insulin infusion pump usage parameters for type-1 diabetic patients under varying meal scenarios and physiological conditions
Research Subject
optimal parameter synthesis and calibration for meal bolus calculation to maintain blood glucose within prescribed ranges and avoid hypoglycemia and hyperglycemia
Publication Details
Publication Date
2012-10-01
Journal
Publisher
ISSN
Cited by
15
Access Type
Author Information
Download PDF
Subscribe to digest
References available in scid.ai4
Meal Simulation Model of the Glucose-Insulin System2007
GIM, Simulation Software of Meal Glucose—Insulin Model2007
Simulation Environment to Evaluate Closed-Loop Insulin Delivery Systems in Type 1 Diabetes2010
A Closed-Loop Artificial Pancreas Using Model Predictive Control and a Sliding Meal Size Estimator2009